Medical Device Recall: Z-1666-2025

FDA Recalls dataset · Firm: Remote Diagnostic Technologies Ltd. Farnborough Aerospace Centre Aerospace Boulevard Farnborough United Kingdom · Status: Open, Classified

Recall Action Class: N/A Date Initiated: 2025-04-16 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-1666-2025.

Recall Res Number
Z-1666-2025
FEI Number
3003832357
Recall Status
Open, Classified
Recall Classification
N/A
Date Initiated
2025-04-16
Date Posted
2025-04-24
Date Terminated
N/A
Product Code
MHX
Event Number
96678
Product Quantity
6,961 units
Product Description Tempus Pro Patient Monitor, REFs: 00-1004-R (Tempus Pro non-printer, IBP, without headset dock), 00-1007-R and 00-1007 (Tempus Pro non-printer, IBP, with headset dock), 00-1024-R (Tempus Pro printer), 00-1026-R (Tempus Pro printer, IBP), 00-1007 [(Tempus Pro, BT Dock (IB Only)], 00-1004 00-1004 [Tempus Pro (IB Only)]

Reason for Recall software issue that causes an error screen on a portable vital signs monitor when attaching video Laryngoscope.

Action Details On April 16, 2025, Remote Diagnostic Technologies Ltd (subsidiary of Philips) issued a Urgent Medical Device Correction to affected consignees via mail. Philips ask consignees to take the following: 1. DO NOT connect the Video Laryngoscope USB probe to the Tempus Pro during device boot-up. Once Patient Details dialog appears, enter patient data, then plug in the Video Laryngoscope USB probe. 2. If the Tempus Pro Monitor displays the error dialog screen (where the user is asked to restart the device), press the Shutdown Button and restart the Tempus Pro Monitor. 3. Keep a copy of this letter with or near your Tempus Pro Monitor. 4. Complete and return the Urgent Field Safety Notification Response Form included with this letter within 30 days of receipt of this notice. 5. This notice must be shared with all relevant personnel within your organization and with any organization where the potentially affected devices have been transferred.

Distribution Pattern Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, WV, WY and the countries of AE, AF, AT, AU, BE, BO, CA, CH, CZ, DE, DK, ES, FI, FR, GB, HK, IE, IL, IS, IT, KY, MY, NL, NO, OM, PL, PT, RO, SA, SE, SG, SI, SL, TH, ZA.

Code Information / Serial Numbers All Lots/ UDI-DI(s): 00-1004-R - 05060472440020 & 05060472442901; 00-1007-R - 05060472440013 & 05060472442918; 00-1024-R - 05060472441027 & 05060472442925; 00-1026-R - 05060472441058 & 05060472442932; 00-1004 - 05060472440624; 00-1007 - 05060472440655

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.