Medical Device Recall: Z-1321-2023

FDA Recalls dataset · Firm: Remote Diagnostic Technologies Ltd. Farnborough Aerospace Centre Aerospace Boulevard Farnborough United Kingdom · Status: Open, Classified

Recall Action Class: N/A Date Initiated: 2023-02-23 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-1321-2023.

Recall Res Number
Z-1321-2023
FEI Number
3003832357
Recall Status
Open, Classified
Recall Classification
N/A
Date Initiated
2023-02-23
Date Posted
2023-04-01
Date Terminated
N/A
Product Code
MHX
Event Number
91805
Product Quantity
5,540 units
Product Description Tempus Pro Patient Monitor is a portable vital signs monitor intended to be used by clinicians and medically qualified personnel, for the attended or unattended monitoring of single or multiple vital signs in clinical and pre hospital care applications. The affected power supplies are related to the following Tempus Pro Patient Monitoring units: Model Numbers: 00-1004 00-1004-R 00-1007 00-1007-R 00-1024-R 00-1026-R Model Numbers: 00-1004 00-1004-R 00-1007 00-1007-R 00-1024-R 00-1026-R

Reason for Recall During internal testing, Regulatory Compliance issues were identified regarding the Monitor and Power Supplies for fluid ingress and basic safety issues.

Action Details On 02/17/2023, the firm sent an "URGENT Medical Device Correction" Letter via priority mail to customers informing them that during internal testing, Regulatory Compliance issues regarding Fluid Ingress and Basic Safety issues were identified with the Tempus Pro Monitor, AC Mains Power Supply, and the Vehicle Adaptor. Customers are instructed to: " As there is currently no indication that any of the issues have resulted in a degradation in product performance, Philips advises that the Tempus Pro devices should remain in service. " Continue to follow the User/Operator Manual, including: o Take care to ensure that water or liquids are not spilt over the device or into its ventilation holes in the side corners. o Check connector covers, particularly the capnometer door cover, to ensure they close and latch acceptably. " Post this Urgent Medical Device Correction letter on or near your Tempus Pro device. " Complete and return the Urgent Medical Device Correction response form included, no later than 30 days from receipt. Philips will be providing a new and compliant AC Mains Power Supply to all affected customers. A new and compliant Vehicle Adaptor will also be provided to affected customers with devices that include that option. Philips anticipates providing this solution to customers in the second half of 2023. For questions or further information, contact the local Philips Representative, or Philip at (800)263-3342.

Distribution Pattern Worldwide Distribution: U.S.(nationwide): AL, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, WV, WY and O.U.S.(foreign): United Arab Emirates, Afghanistan, Austria, Australia, Belgium, Brunei Darussalam, Bolivia, Canada, Switzerland, Czechia, Germany, Denmark, Spain, France, United Kingdom, Hong Kong, Ireland, Israel, Iceland, Italy, Cayman Islands, Luxembourg, Malaysia, Netherlands, Norway, Portugal, Romania, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Mayotte, and South Africa.

Code Information / Serial Numbers Model Numbers: 00-1004 00-1004-R UDI-DI Codes: 05060472440020 05060472442901 00-1007 00-1007-R UDI-DI Codes: 05060472440013 05060472442901 00-1024-R UDI-DI Codes: 05060472441027 05060472442925 00-1026-R UDI-DI Codes: 05060472441058 05060472442932 Lot Codes - All products manufactured are affected.

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.