Product Description
Arietta 850 Ultrasound-
Software Version: V1.0.0 through V4.1.3
Reason for Recall
Diagnostic ultrasound system with the specified
software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display
the measurement results correctly with two cardiology measurement functions:
Mitral Regurgitation Flow (MR Flow) Measurement;
Mitral Valve (MV) Measurement.
In addition there is an error in the instruction manual.
Action Details
Fujifi Film issued Medical Device Correction Notice letter dated 10/07/21 via courier on 10/14/21.
Letter states reason for recall, health risk and action to take:
Customers/Users are advised to accept the corrected software and manual as soon as available.
Advisory Notification applies to all customers and distributors of the ARIETTA 850 systems with software version V1.0.0 through V4.1.3, See following page for complete list of affected systems.
We will contact you to set up time to update the software and provide the revised manual. (7) Other Information:
For questions, comments, or concerns, please contact:
Aaron Pierce
Director, QA/RA
1959 Summit Commerce Park
Twinsburg, OH 44087
Phone (330) 425-1313 x2934, Fax (330) 963-0749
Email aaron.pierce@fujifilm.com
Distribution Pattern
Nationwide
Code Information / Serial Numbers
Serial Numbers:
205W3386 (01)04562122958974(11)171114(21)205W3386;
206F7900 (01)04562122958974(11)180926(21)206F7900;
206F7901 (01)04562122958974(11)180926(21)206F7901;
206F7936 (01)04562122958974(11)181113(21)206F7936;
206F7937 (01)04562122958974(11)181113(21)206F7937;
206F7940 (01)04562122958974(11)181113(21)206F7940;
206F7941 (01)04562122958974(11)181113(21)206F7941;
206F7942 (01)04562122958974(11)181113(21)206F7942;
206F7944 (01)04562122958974(11)181113(21)206F7944;
206F7945 (01)04562122958974(11)181113(21)206F7945;
206F7946 (01)04562122958974(11)181113(21)206F7946;
206F7947 (01)04562122958974(11)181114(21)206F7947;
G3002552 (01)04562122958974(11)20190115(21)G3002552;
G3024545 (01)04562122958974(11)190607(21)G3024545;
G3037885 (01)04562122958974(11)20190719(21)G3037885;
G3037887 (01)04562122958974(11)20190719(21)G3037887;
G3042405 (01)04562122958974(11)190829(21)G3042405;
G3042823 (01)04562122958974(11)20190807(21)G3042823;
G3042828 (01)04562122958974(11)20190808(21)G3042828;
G3045094 (01)04562122958974(11)20190822(21)G3045094;
G3052105 (01)04562122958974(11)20191008(21)G3052105
G3052118 (01)04562122958974(11)20191025(21)G3052118
G3055549 (01)04562122958974(11)20191118(21)G3055549
G3055550 (01)04562122958974(11)20191118(21)G3055550
G3055552 (01)04562122958974(11)20191118(21)G3055552
G3055553 (01)04562122958974(11)20191118(21)G3055553
G3055554 (01)04562122958974(11)20191118(21)G3055554
G3055555 (01)04562122958974(11)20191118(21)G3055555
G3060234 (01)04562122958974(11)20191206(21)G3060234
G3060254 (01)04562122958974(11)20191206(21)G3060254
G3060275 (01)04562122958974(11)20191216(21)G3060275
G3060278 (01)04562122958974(11)20191216(21)G3060278
G3060690 (01)04562122958974(11)200127(21)G3060690
G3071074 (01)04562122958974(11)20200227(21)G3071074
G3080785 (01)04562122958974(11)201105(21)G3080785
G3090450 (01)04562122958974(11)201225(21)G3090450
G3110884 (01)04562122958974(11)210401(21)G3110884
Public Record Disclaimer
This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.