Medical Device Recall: Z-0034-2019

FDA Recalls dataset · Firm: Oridion Medical 1987 Ltd. Har Hotzvim Industrial · Status: Terminated

Recall Action Class: N/A Date Initiated: 2018-07-31 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0034-2019.

Recall Res Number
Z-0034-2019
Recalling Firm / Manufacturer
FEI Number
3001227743
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2018-07-31
Date Posted
N/A
Date Terminated
2022-09-20
Product Code
DQA
Event Number
80895
Product Quantity
48
Product Description Capnostream20 (US) N W/PRINTER REFURBISHED, Part Number CS78657

Reason for Recall The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory default settings when the monitor is powered off.

Action Details Urgent Medical Device Correction notification letters dated 7/31/18 were distributed to customers. The letters included the following instructions for customers: Share this notification with all care environments where the Capnostream" 20 and Capnostream" 20p bedside patient monitors are used, particularly in the areas where nurse call may be enabled (general care floors). If your facility has distributed these bedside patient monitors to other persons or facilities, please promptly forward a copy of this letter to those recipients. Medtronic is recommending continued use of Capnostream" devices. Users should confirm that the date/time is accurately displayed. If the date/time requires reset, all user defined settings, such as alarms and nurse call, will also require reset until the software is updated. Please follow the instruction in the Operators Manual to set user-defined default settings. The Operators Manual is available at http://www.medtronic.com/covidien/enus/support/product-manuals.html, refer to the Institutional Settings section. Return the completed Acknowledgement and Receipt Form by fax or email even if you have no inventory. In October 2018, download the software update from the Medtronic website via the following link which includes directions on how to download and install the software: http://www.medtronic.com/covidien/en-us/support/patient-monitoring-equipmentsoftware-upgrades.html.

Distribution Pattern The products were distributed US nationwide and Puerto Rico. The products were distributed to the following foreign countries: Argentina, Australia, Austria, Bahrain, Belgium, Canada, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Kazakhstan, Kuwait, Mexico, Netherlands, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, UAE, UK, and Vietnam.

Code Information / Serial Numbers B200001412 B200001550 B200001579 B200001485 B200002038 B200001062 B200001148 B200001184 B200001273 B200001303 B200001320 B200001332 B200001351 B200001372 B200001384 B200001444 B200001083 B200001394 B200001552 B200001200 B200001233 B200001553 B200001121 B200001123 B200001446 B200001036 B200001124 B200001125 B200001130 B200001035 B200001112 B200001482 B200001095 B400002305 B200001143 B200001252 B400002682 B200001064 B200001435 B200001489 B400002288 B200001136 B200001355 B200001608 B200001998 B200001349 B200001066 B410123252

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.