Medical Device Recall: Z-0001-2015

FDA Recalls dataset · Firm: Invivo Corporation · Status: Terminated

Recall Action Class: N/A Date Initiated: 2014-09-08 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0001-2015.

Recall Res Number
Z-0001-2015
Recalling Firm / Manufacturer
FEI Number
1051786
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2014-09-08
Date Posted
2014-10-01
Date Terminated
2017-01-09
Product Code
MWI
Event Number
69192
Product Quantity
10 units
Product Description Philips Expression MR200 MRI Patient Monitoring System Model 866120

Reason for Recall During production, excess metal (aluminum) debris was found stuck to the cling plastic that covers the device's pole assembly. The metallic debris could affect system performance due to the potential for a failure to power up or loss of power. This could result in the loss of monitoring and a delay in diagnosis of the patient.

Action Details The firm, Philips Healthcare, sent an "URGENT- Medical Device Recall" letter dated June 17, 2014, to its customers informing them of the product, problem and actions taken. The firm informed the customers that there are no additional actions required on their part. The Philips/Invivo Account Manager will contact them in order to arrange for replacement of the device. If you need any further information or support concerning this issue, contact your local Philips/Invivo representative: 877-468-4861 option 1, then option 2.

Distribution Pattern Worldwide Distribution: US distribution in state of: CA and countries of: Poland, Singapore, and Sweden.

Code Information / Serial Numbers Model 866120 with the following lots #'s: US33600006, US33600007, US33600011, US33600012 and US33600013

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.

Account Required Full available specs will require a free AvaSpecs account.