Product Description
Synapse CV 6. with AR.
A web-based application as the primary user interface for the processing of medical images.
Reason for Recall
The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.
Action Details
On January 3, 2024, the firm notified customers via URGENT MEDICAL DEVICE CORRECTION letters.
Fujifilm will offer to correct the formula that calculates the LV Mass (2D Bullet) in the Synapse database. Until the update is scheduled and completed, customers may continue to safely use the Pediatric Workflow modules listed in the notice as long as the LV Mass (2D Bullet) equation is not the main factor for a diagnosis.
Customers should coordinate with the Fujifilm service engineer to schedule the correction of their device when contacted.
Please contact Fujifilm's field action coordinator, Kotei Aoki, via phone at 765-246-2931 or email at kotei.aoki@fujifilm.com, if you have any further questions regarding this field action.
Distribution Pattern
US domestic distribution nationwide. International distribution to Israel.
Code Information / Serial Numbers
UDI-DI 00854904006015
Software versions 6.2.1, 6.2.0, 6.1.1, 6.1.0, 6.0.1, and 6.0.0
Public Record Disclaimer
This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.