Medical Device Recall: Z-0139-2011

FDA Recalls dataset · Firm: Siemens Healthcare Diagnostics, Inc. · Status: Terminated

Recall Action Class: N/A Date Initiated: 2010-08-19 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0139-2011.

Recall Res Number
Z-0139-2011
Recalling Firm / Manufacturer
FEI Number
2517506
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2010-08-19
Date Posted
2010-10-25
Date Terminated
2010-10-26
Product Code
CEM
Event Number
56703
Product Quantity
3,706 cartons
Product Description Dimension Vista V-LYTE Standard A, Catalog No K820 For the calibration of Na+/K+/Cl- on the Dimension vista System.

Reason for Recall High potassium content in the affected lots results in the potential for low potassium recovery for Dilution Check, QC and patient samples when run on the Dimension Vista System.

Action Details Siemens sent an Urgent Field Safety Notice (dated "August 2010") to customers identifying the product, the problem and the action to be taken. Customers were instructed on how to continue testing until new lots of the recalled product were available and asked to complete and fax a Field Correction Effectiveness Check to 301-631-8467 to confirm receipt of the Urgent Field Safety Notice. For technical questions or concerns customers were instructed to contact the Siemens Technical Solutions Center at 800-441-9250.

Distribution Pattern USA, Nationwide Distribution

Code Information / Serial Numbers KA0C03 EXP 3/24/11, KA0D01 EXP 4/14/11, KA0E01 EXP 5/5/11, KA0E02 EXP 5/19/11, KA0F01 EXP 6/3/11, AND KA0G01 EXP 7/8/11

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.