Medical Device Recall: Z-0112-2015

FDA Recalls dataset · Firm: Nellcor Puritan Bennett Inc. (dba Covidien LP) · Status: Terminated

Recall Action Class: N/A Date Initiated: 2014-10-01 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0112-2015.

Recall Res Number
Z-0112-2015
Recalling Firm / Manufacturer
FEI Number
2936999
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2014-10-01
Date Posted
2014-10-28
Date Terminated
2016-05-02
Product Code
CBK
Event Number
69320
Product Quantity
324 units
Product Description Covidien Puritan Bennett 980 Ventilator, Rx ONLY. Suitable for service in a hospital (institutions) and intra-hospital transport to provide continuous positive pressure ventilator for Neonatal (NICU) through Adult patient populations.

Reason for Recall A software issue may lead to ventilator inoperative situations.

Action Details The firm sent Urgent: Field Correction Action letters, dated October 3, 2014, to consignees. The letter identified the affected product and the reason for the recall. Customers were provided actions to be taken, as well as, important safety reminders. The letter informed customers that they may continue to use their ventilators pending the software correction as long as two gas sources are connected to the ventilator at all times. Customers are to complete the attached acknowledgement and receipt form and fax it to Covidien at the number provided. For further assistance, customers are to contact the Technical Support Department at 1-800-255-6774.

Distribution Pattern Worldwide Distribution -- USA, including the states of CA, CO, FL, GA, KY, MA, MN, NC, NY, OH, OK, PA, SC, TN, TX, UT, WI; and, the countries of Canada, Mexico, Saudi Arabia, South Africa, and United Arab Emirates.

Code Information / Serial Numbers Model number PB980 Ventilator:35B1400143, 35B1400684, 35B1400627, 35B1400708, 35B1400398, 35B1400482,  35B1400491, 35B1400512, 35B1400601, 35B1400121, 35B1400568, 35B1400590, 35B1400595, 35B1400599, 35B1400606, 35B1400106, 35B1400146, 35B1400169, 35B1400226, 35B1400260, 35B1400270, 35B1400263, 35B1400351, 35B1400373, 35B1400400, 35B1400401, 35B1400608, 35B1400089, 35B1400112, 35B1400683, 35B1400685, 35B1400678, 35B1400118, 35B1400114, 35B1400355, 35B1400144, 35B1400300, 35B1400379, 35B1400412, 35B1400472, 35B1400489, 35B1400490, 35B1400492, 35B1400493, 35B1400499, 35B1400500, 35B1400501, 35B1400502, 35B1400507, 35B1400577,  35B1400207, 35B1400211, 35B1400216, 35B1400222, 35B1400225, 35B1400231, 35B1400267, 35B1400312, 35B1400323, 35B1400340, 35B1400383, 35B1400128, 35B1400363, 35B1400459, 35B1400589, 35B1400616,  35B1400637, 35B1400483, 35B1400311, 35B1400447,  35B1400648, 35B1400651, 35B1400654, 35B1400662,  35B1400686, 35B1400692, 35B1400175, 35B1400294, 35B1400436, 35B1400485, 35B1400586, 35B1400644,  35B1400687, 35B1400689, 35B1400694, 35B1400700,  35B1400661, 35B1400676, 35B1400693, 35B1400732,  35B1400734, 35B1400736, 35B1400737, 35B1400742, 35B1400280, 35B1400360, 35B1400471, 35B1400609,  35B1400674, 35B1400688, 35B1400755, 35B1400480,  35B1400675, 35B1400706, 35B1400771, 35B1400667,  35B1400704, 35B1400705, 35B1400727, 35B1400743,  35B1400279, 35B1400285, 35B1400262, 35B1400265, 35B1400259, 35B1400494, 35B1400403, 35B1400779, 35B1400783, 35B1400509, 35B1400724, 35B1400763, 35B1400767, 35B1400566, 35B1400768, 35B1400772,  35B1400791, 35B1400793, 35B1400817, 35B1400525, 35B1400796, 35B1400801, 35B1400809, 35B1400821,  35B1400828, 35B1400333, 35B1400366, 35B1400392,  35B1400440, 35B1400463, 35B1400478, 35B1400308,  35B1400347, 35B1400367, 35B1400396, 35B1400078,  35B1400099, 35B1400130, 35B1400155, 35B1400157, 35B1400158, 35B1400174, 35B1400176, 35B1400178, 35B1400180, 35B1400281, 35B1400315, 35B1400631, 35B1400635, 35B1400625, 35B1400653, 35B1400843, 35B1400847, 35B1400233, 35B1400271, 35B1400337, 35B1400350, 35B1400359, 35B1400368, 35B1400384, 35B1400468, 35B1400486, 35B1400495, 35B1400094, 35B1400236, 35B1400437, 35B1400438, 35B1400479, 35B1400496, 35B1400505, 35B1400506, 35B1400427, 35B1400511, 35B1400325, 35B1400462, 35B1400514, 35B1400536, 35B1400537, 35B1400544, 35B1400066, 35B1400200, 35B1400782, 35B1400829, 35B1400015, 35B1400110, 35B1400516, 35B1400517, 35B1400859, 35B1400013, 35B1400103, 35B1400167, 35B1400124, 35B1400007, 35B1400058, 35B1400083, 35B1400129, 35B1400132, 35B1400006, 35B1400079, 35B1400113, 35B1400116, 35B1400219, 35B1400163, and 35B1400543.

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.