Medical Device Recall: Z-0038-2012

FDA Recalls dataset · Firm: Siemens Healthcare Diagnostics, Inc. · Status: Terminated

Recall Action Class: N/A Date Initiated: 2011-08-31 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0038-2012.

Recall Res Number
Z-0038-2012
Recalling Firm / Manufacturer
FEI Number
2517506
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2011-08-31
Date Posted
2011-10-12
Date Terminated
2013-05-20
Product Code
JJE
Event Number
59888
Product Quantity
998
Product Description Dimension(R) EXL LOCI Cardiac Troponin I Calibrator (RC621) The LOCI Cardiac TNI Calibrator is an in vitro diagnostic product for the calibration on Cardiac Troponin I method on the Dimension(R) EXL (TM) Integrated chemistry system with LOCI(R) module.

Reason for Recall Internal testing has confirmed that recovery of levels 2 and 3 of LOCl Cardiac Troponin I Calibrator is lower than expected bottle value. Following calibration with LOCl Cardiac Troponin I Calibrator, lot 1AD045 an upward shift in patient and QC results may be observed.

Action Details Siemens sent an Urgent Field Safety Notice dated August 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue use and discard any remaining inventory. Replacement orders would be processed and sent free of charge. Customers were also instructed to complete the attached FIELD CORRECTION EFFECTIVENESS CHECK form and fax it to (302) 631- 8457. Customers were instructed to forward the Safety Notice to anyone to whom they may have distributed the affected product. For any technical questions customers were to contact the Siemens Solutions Center at 800-441-9250.

Distribution Pattern Worldwide Distribution - USA (nationwide) including AL, AK, AZ, AR, CA, CO, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY and the countries of GUAM, Austria, Canada, China, Czech Republic, Egypt, France, Germany, Italy, Poland, Portugal, Puerto Rico, Spain and Switzerland.

Code Information / Serial Numbers Lot Number 1AD045

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.

Account Required Full available specs will require a free AvaSpecs account.