Medical Device Recall: Z-0011-2018

FDA Recalls dataset · Firm: Abbott Laboratories, Inc · Status: Terminated

Recall Action Class: N/A Date Initiated: 2015-01-15 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0011-2018.

Recall Res Number
Z-0011-2018
Recalling Firm / Manufacturer
FEI Number
1628664
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2015-01-15
Date Posted
N/A
Date Terminated
2020-07-20
Product Code
JJE
Event Number
77828
Product Quantity
1,385 cuvette segments
Product Description ARCHITECT c4000 Cuvette Segment, List number 02P75-01. The firm name on the label is Abbott Laboratories, Abbott Park, IL. Cuvette segments are a component to the ARCHITECT c4000 system. Cuvette segments are racks that sit in the reaction carousel and hold cuvettes. Each cuvette segment holds 11 cuvettes. With 9 segments, the reaction carousel holds 99 cuvettes. The Abbott ARCHITECT System is intended for In Vitro diagnostic use only. The system is designed to perform automated chemistry tests, utilizing photometry and potentiometric technology.

Reason for Recall There is a potential to generate falsely-depressed patient results in the cuvettes adjacent to the detached segment post for any assay run on the c4000 instrument.

Action Details The firm, Abbott Laboratories, sent a "Product Correction" letter dated June 26, 2017 to is customers via overnight mail who had not had the Field Action Mandatory Technical Service Bulletin completed on their instrument. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: -Contact their representative to schedule an inspection of their cuvette segments. -If a detached segment is identified o A replacement segment will be provided at no charge. o Follow your laboratory protocol regarding previously reported patient results from cuvettes within the impacted segment. -If you have forwarded the product listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of the letter. -Retain the letter for your laboratory records. -Complete and return the Customer Reply Form, prior to July 11, 2017 via FAX# 1-800-777-0051 or email QAGCO@abbott.com. Even if you no longer have the instrument(s). Note: consignees who have the instruments that have not had the FA Mandatory TSB completed are the ones who are targeted for this current field action. If you or any of the health care providers you serve have any questions regarding this information, U.S. Customers please contact Customer Service at 18774ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., please contact your local area Customer Service.

Distribution Pattern Worldwide Distribution: US (nationwide) including Puerto Rico, and countries of: Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bangladesh, Belarus, Belgium, Bosnia & Herzegovina, Brazil, Brunei, Bulgaria Burkina Faso, Canada, Chile, China, Columbia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Gaza & Jericho, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Ivory Coast, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malawi, Malaysia, Maldives, Mali, Mauritania, Mexico, Morocco, Netherlands, New Zealand, Nigeria, North Yemen, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia (CIS), Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, St. Lucia, Sweden, Switzerland, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turkey, Turks & Caicos, UAE, Uganda, United Kingdom, Uruguay, Venezuela, Viet Nam, Zambia, and Zimbabwe.

Code Information / Serial Numbers All c4000s

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.