Medical Device Recall: Z-0011-2014

FDA Recalls dataset · Firm: Siemens Healthcare Diagnostics · Status: Terminated

Recall Action Class: N/A Date Initiated: 2013-08-12 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0011-2014.

Recall Res Number
Z-0011-2014
Recalling Firm / Manufacturer
FEI Number
2432235
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2013-08-12
Date Posted
2013-10-17
Date Terminated
2015-11-18
Product Code
GKZ
Event Number
65898
Product Quantity
311 units total (US: 51 units; Foreign: 260 units)
Product Description ADVIA 2120/2120i Hematology Systems with ADVIA Autoslide connected to it, SMN 10361162 and SMN 10361798 --- CLASSIFICATION NAME: Counter, Differential Cell, Automated (Particle Counter) The ADVIA 2120/2120i are hematology systems that utilizes the principles of flow cytometry in order to provide complete blood counts.

Reason for Recall Siemens determined that if an autosampler rack jam error occurs during operation of an ADVIA 2120/2120i connected to an ADVIA Autoslide, it is possible that the next slide processed by the Autoslide could be labeled with the wrong sample identification information.

Action Details Siemens sent an Urgent Medical Device Correction Letter dated August 8, 2013, and Effectiveness Checks Forms were sent to customers on August 12, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were advised if they do not have an ADVIA Autoslide connected to their system, no action is required. To avoid the possibility of mislabeling a slide when an ADVIA 2120/2120i rack jam error occurs: Eject the sample rack. Troubleshoot and correct the reason for the rack jam error. Rerun the sample rack. Field service personnel were sent a support bulletin describing the issue and instructing them how to deal with customer questions. If you need additional assistance, please contact Global Product Support Tarrytown, New York @ osggps.hschematology.healthcare@siemens.com

Distribution Pattern Worldwide Distribution - USA (nationwide) and the countries of Australia, Austria, Belgium, Canada, Croatia, Denmark, France, Germany, Greece, Italy, Japan, Malaysia, Netherlands, Norway, Philippines, Portugal, South Korea, Singapore, Slovenia, Spain, Switzerland, Thailand, Turkey, United Kingdom and Vietnam.

Code Information / Serial Numbers All serial numbers connected to the ADVIA 2120/2120i Hematology Systems with ADVIA Autoslide.

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.