St Jude Medical Inc.

Entity profile compiled from public source records

Public Record Profile Registrations: 0 510(k) filings: 0 Recalls: 15

Known Identifiers & Locations

FEI Numbers 1627487, 2017865
Owner Operator Numbers N/A
Locations Covered N/A

Registered Establishments (0)

No device registration and listing records found for this entity.

510(k) Premarket Notifications (0)

No 510(k) notifications found for this entity.

FDA Recall History (15)

Recall # Product Description Initiation Date Status Action
Z-0003-2018 Fortify VR - Model Numbers: CD1231-40 and CD1231-40Q; Fortify ST VR, Model Numbers: CD1241-40 and CD1241-40Q; Fortify Assura VR, Model Numbers: CD1257-40, CD1257-40Q, CD1357-40C, CD1357-40Q; Fortify Assura ST VR, Model Numbers: CD1263-40, CD1263-40Q, CD1363-40C, CD1363-40Q; Fortify DR, Model Numbers: CD2231-40 and CD2231-40Q; Fortify ST DR, Model Numbers: CD2241-40 and CD2241-40Q; Fortify Assura DR, Model Numbers: CD2257-40, CD2257-40Q, CD2357-40C, CD2357-40Q; Fortify Assura ST DR, Model Numbers: CD2263-40, CD2263-40Q, CD2363-40C, CD2363-40Q Product Usage Fortify and Fortify Assura implantable cardioverter defibrillators (ICDS) are intended to provided ventricular anti-tachycardia pacing and ventricular defibrillation for automated treatment of life threatening ventricular arrhythmias. The devices may also be programmed to provide single or dual chamber bradycardia pacing Reason: The device may exhibit premature battery depletion. 2017-08-28 Terminated View Details
Z-0004-2018 Unify, Model Numbers: CD3231-40 and CD3231-40Q; Unify Quadra, Model Numbers: CD3249-40 and CD3249-40Q; Unify Assura, Model Numbers: CD3257-40, CD3257-40Q, CD3357-40C, CD3357-40Q; Quadra Assura, Model Numbers: CD3265-40, CD3265-40Q, CD3365-40C, CD3365-40Q; Quadra Assura MP, Model Numbers: CD3269-40, CD3269-40Q, CD3369-40C, CD3369-40Q Unify and Unify Assura cardiac resynchronization therapy defibrillators (CRTDs) are intended to provide ventricular anti-tachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias. The devices may also be programmed to provide single or dual chamber bradycardia pacing, in additional to cardiac resynchronization therapy via left ventricular pacing Reason: The device may exhibit premature battery depletion. 2017-08-28 Terminated View Details
Z-0022-2018 Proclaim 5 Elite Implantable Pulse Generator and Proclaim 7 Elite Implantable Pulse Generator Models 3660, 3661, 3662, 3663, 3665, 3667 Reason: Analysis of the difference between the actual device longevity and the battery indicator status provided by the Clinician Programmer (CP) or Patient Controller (PC) has revealed errors in the CP or PC longevity estimations that can overestimate or under estimate the battery indicator status in certain situations. 2017-09-12 Terminated View Details
Z-0023-2018 Infinity 5 Implantable Pulse Generator and Infinity 7 Implantable Pulse Generator Models 6660, 6661, 6662, 6663 Reason: Analysis of the difference between the actual device longevity and the battery indicator status provided by the Clinician Programmer (CP) or Patient Controller (PC) has revealed errors in the CP or PC longevity estimations that can overestimate or under estimate the battery indicator status in certain situations. 2017-09-12 Terminated View Details
Z-0024-2018 Proclaim DRG Implantable Pulse Generator Model 3664 Reason: Analysis of the difference between the actual device longevity and the battery indicator status provided by the Clinician Programmer (CP) or Patient Controller (PC) has revealed errors in the CP or PC longevity estimations that can overestimate or under estimate the battery indicator status in certain situations. 2017-09-12 Terminated View Details
Z-0029-2018 Anthem Model Numbers: PM3210, PM3212; Allure/Relieve Model Numbers: PM3222, PM3242, PM3262, PM3542, PM3224, PM3244 Implantation of a single-chamber pulse generator, dual-chamber pulse generator, or CRT-P is indicated in one or more of the following permanent conditions: Syncope, Presyncope, Fatigue, Disorientation, Or any combination of those symptoms. Implantation of a CRT-P is indicated for: Maintaining synchrony of the left and right ventricles in patients who have undergone an AV nodal ablation for chronic atrial fibrillation and have NYHA Class II or III heart failure. The reduction of the symptoms of moderate to severe heart failure (NYHA Class III or IV) in those patients who remain symptomatic despite stable, optimal medical therapy, and have a left ventricular ejection fraction d35% and a prolonged QRS duration Reason: New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications. 2017-08-28 Terminated View Details
Z-0030-2018 Assurity Model Numbers: PM1240, PM2240; Assurity + Model Number: PM2260; Accent Model Numbers: PM1210, PM2210, PM1222, PM2212; Accent MRI Model Numbers: PM1224, PM2218, PM2224 Implantation of a single-chamber pulse generator, dual-chamber pulse generator, or CRT-P is indicated in one or more of the following permanent conditions: Syncope, Presyncope, Fatigue, Disorientation, Or any combination of those symptoms. Implantation of a CRT-P is indicated for: Maintaining synchrony of the left and right ventricles in patients who have undergone an AV nodal ablation for chronic atrial fibrillation and have NYHA Class II or III heart failure. The reduction of the symptoms of moderate to severe heart failure (NYHA Class III or IV) in those patients who remain symptomatic despite stable, optimal medical therapy, and have a left ventricular ejection fraction d35% and a prolonged QRS duration Reason: New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications. 2017-08-28 Terminated View Details
Z-0031-2018 Assurity MRI Model Numbers: PM1272, PM2272 Implantation of a single-chamber pulse generator, dual-chamber pulse generator, or CRT-P is indicated in one or more of the following permanent conditions: Syncope, Presyncope, Fatigue, Disorientation, Or any combination of those symptoms. Implantation of a CRT-P is indicated for: Maintaining synchrony of the left and right ventricles in patients who have undergone an AV nodal ablation for chronic atrial fibrillation and have NYHA Class II or III heart failure. The reduction of the symptoms of moderate to severe heart failure (NYHA Class III or IV) in those patients who remain symptomatic despite stable, optimal medical therapy, and have a left ventricular ejection fraction d35% and a prolonged QRS duration Reason: New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications. 2017-08-28 Terminated View Details
Z-0032-2018 Merlin PCS 3650 Programmer with Model 3330 Software Version 23.1.1 The Merlin Patient Care System (PCS) Model 3650 Programmer is a portable, dedicated programming system designed to interrogate program display data and test St. Jude Medical implantable devices. Reason: New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications. 2017-08-28 Terminated View Details
Z-0033-2018 Merlin@Home Software Model EX2000 v8.2.2 for Merlin@Home Transmitter Models EX1150, EX1150W, EX1100, and EX1100W. The Merlin@home Transmitters are a component of St. Jude Medicals (SJM) Remote Care System. These Merlin@home transmitters are used to retrieve information from the SJM implantable medical device (IMD) using RF telemetry. The data is then sent back to the Merlin.net Patient Care Network through a built-in modem or, alternatively, through a USB cellular adapter or a USB based Broadband Kit. The transmitted data enables clinicians to remotely monitor a patients IMD using a secure, encrypted, password-protected web site, thus reducing the number of in-clinic visits a patient needs to make to their physician. The transmitter monitors the patients device daily between scheduled follow-ups and can alert a clinician if it detects anything out of the norm. Reason: New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications. 2017-08-28 Terminated View Details
Z-0034-2018 PM1214 NUANCE SR RF; PM1230 and PM2230 NUANCE MRI; PM2214 NUANCE DR RF These low voltage (LV) devices are implantable pacemaker pulse generators, intended to be permanently implanted in the body, that have a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. These devices are used as a substitute for the heart's intrinsic pacing system to correct both intermittent and continuous cardiac rhythm disorders. Reason: New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications. 2017-08-28 Terminated View Details
Z-0035-2018 PM1226 ACCENT ST MRI SR RF and PM2222 ACCENT ST These low voltage (LV) devices are implantable pacemaker pulse generators, intended to be permanently implanted in the body, that have a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. These devices are used as a substitute for the heart's intrinsic pacing system to correct both intermittent and continuous cardiac rhythm disorders. Reason: New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications. 2017-08-28 Terminated View Details
Z-0036-2018 PM 1250 and PM 2250 ZENEX; PM1282 and PM2282 ZENEX MRI These low voltage (LV) devices are implantable pacemaker pulse generators, intended to be permanently implanted in the body, that have a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. These devices are used as a substitute for the heart's intrinsic pacing system to correct both intermittent and continuous cardiac rhythm disorders. Reason: New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications. 2017-08-28 Terminated View Details
Z-0037-2018 PM 1260 ASSURITY + These low voltage (LV) devices are implantable pacemaker pulse generators, intended to be permanently implanted in the body, that have a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. These devices are used as a substitute for the heart's intrinsic pacing system to correct both intermittent and continuous cardiac rhythm disorders. Reason: New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications. 2017-08-28 Terminated View Details
Z-0038-2018 PM3544 QUADRA RELIEVE and PM3562 QUADRA ALLURE MP These low voltage (LV) devices are implantable pacemaker pulse generators, intended to be permanently implanted in the body, that have a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. These devices are used as a substitute for the heart's intrinsic pacing system to correct both intermittent and continuous cardiac rhythm disorders Reason: New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications. 2017-08-28 Terminated View Details