Medical Device Recall: Z-0033-2018

FDA Recalls dataset · Firm: St Jude Medical Inc. · Status: Terminated

Recall Action Class: N/A Date Initiated: 2017-08-28 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0033-2018.

Recall Res Number
Z-0033-2018
Recalling Firm / Manufacturer
FEI Number
2017865
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2017-08-28
Date Posted
N/A
Date Terminated
2020-07-08
Product Code
OSR
Event Number
78093
Product Quantity
495,011 units (391,093 in US)
Product Description Merlin@Home Software Model EX2000 v8.2.2 for Merlin@Home Transmitter Models EX1150, EX1150W, EX1100, and EX1100W. The Merlin@home Transmitters are a component of St. Jude Medicals (SJM) Remote Care System. These Merlin@home transmitters are used to retrieve information from the SJM implantable medical device (IMD) using RF telemetry. The data is then sent back to the Merlin.net Patient Care Network through a built-in modem or, alternatively, through a USB cellular adapter or a USB based Broadband Kit. The transmitted data enables clinicians to remotely monitor a patients IMD using a secure, encrypted, password-protected web site, thus reducing the number of in-clinic visits a patient needs to make to their physician. The transmitter monitors the patients device daily between scheduled follow-ups and can alert a clinician if it detects anything out of the norm.

Reason for Recall New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.

Action Details Abbott sent an Important Cybersecurity Advisory dated August 28, 2017, to all affected customers to notify customers of the availability of the programmer software update and associated pacemaker firmware update. The notification material is in the form of a Physician Letter and Hospital Letter and will be delivered to physicians and hospitals with inventory in the US by overnight service. Customers with questions were instructed to contact their Abbott representative or the customers technical support hotline at 1-800-722-3774.

Distribution Pattern Worldwide Distribution

Code Information / Serial Numbers Model EX1100 - GTIN 05414734504782 and Model EX1150 - GTIN 05414734404334

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.