Product Description
Applied Medical, Separator Abdominal Access System REF: C0606 15x150mm Non-threaded Separator System with Universal Seal, for laparoscopic procedures.
Reason for Recall
Potential inability to insufflate through the stopcock. Under certain circumstances, a flexible elastomeric component inside the 15mm trocar can
stretch into a configuration that blocks the flow of insufflation gas (C02).
Action Details
Customers were notified via FedEx or UPS on June 25,2008 with an overnight letter. The letter advises that Applied Medical is conducting a voluntary recall of the 15mm Separator Access Systems, Models C0604, C0605, C0606 and C0607 due to the potential inability to insufflate through the stopcock. Per the letter, the firm is asking
that all 15mm product listed in the letter be returned. A response form is provided.
Distribution Pattern
Nationwide and Canada distribution.
Code Information / Serial Numbers
Product/Kit Batch Number 1059804
Public Record Disclaimer
This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.