Medical Device Recall: Z-0016-2025

FDA Recalls dataset · Firm: Smiths Medical ASD Inc. · Status: Open, Classified

Recall Action Class: N/A Date Initiated: 2024-06-26 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0016-2025.

Recall Res Number
Z-0016-2025
Recalling Firm / Manufacturer
FEI Number
3012307300
Recall Status
Open, Classified
Recall Classification
N/A
Date Initiated
2024-06-26
Date Posted
2024-10-18
Date Terminated
N/A
Product Code
FPA
Event Number
95305
Product Quantity
8976 units
Product Description CADD Medication Cassette Reservoir, 50mL, FS, Product Code 21-7301-24JP

Reason for Recall Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.

Action Details Smiths Medical issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 06/26/2024 via mail. The notice explained the issue, potential risks, and requested the affected product be quarantined, and the response for returned to initiate the return and replacement process. Distributors were directed to notify their customers. For further inquires: Product complaints or adverse events: productcomplaints@icumed.com 1-(866)-216-8806 Additional information or technical assistance 1-(800)-241-4002, option 3 tsc.support@icumed.com

Distribution Pattern Worldwide distribution.

Code Information / Serial Numbers UDI/DI NA, Lot/Serial Numbers: 4013366, 4034019, 4037749, 4034021, 4040248, 4042841, 4048891, 4052407

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.