| Z-0004-2012 |
Smiths Medical, CADD-Solis ambulatory infusion pumps, Models 2100, REF 21-2101-0200-51 and 2110, REF 21-2111-0100-51, Rx Only.
Indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site (soft tissue, body cavity/ surgical wound site), epidural space, or subarachnoid space infusion.
Reason: Reports of non-delivery/ significant under-delivery of fluids when the CADD Medication Cassette Reservoir is not correctly latched to the pump.
EXPANDED: On 7/14/11, Smith Medical expanded this recall to include all consignees with affected pumps.
|
2011-05-10 |
Terminated
|
View Details
|
| Z-0010-2025 |
CADD Medication Cassette Reservoir, 100mL, Product Code 21-7002-24
Reason: Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
|
2024-06-26 |
Open, Classified
|
View Details
|
| Z-0011-2025 |
CADD Medication Cassette Reservoir, 100mL, FS, Product Code 21-7302-24
Reason: Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
|
2024-06-26 |
Open, Classified
|
View Details
|
| Z-0012-2025 |
CADD Medication Cassette Reservoir, 100mL, FS, YELLOW, Product Code 21-7300-24
Reason: Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
|
2024-06-26 |
Open, Classified
|
View Details
|
| Z-0013-2025 |
CADD Medication Cassette Reservoir, 50mL, FS, Product Code 21-7301-24
Reason: Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
|
2024-06-26 |
Open, Classified
|
View Details
|
| Z-0014-2025 |
CADD Medication Cassette Reservoir, 50mL, Product Code 21-7001-24
Reason: Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
|
2024-06-26 |
Open, Classified
|
View Details
|
| Z-0015-2025 |
CADD Medication Cassette Reservoir, 100mL, FS, Product Code 21-7302-24JP
Reason: Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
|
2024-06-26 |
Open, Classified
|
View Details
|
| Z-0016-2025 |
CADD Medication Cassette Reservoir, 50mL, FS, Product Code 21-7301-24JP
Reason: Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
|
2024-06-26 |
Open, Classified
|
View Details
|
| Z-0017-2025 |
ENT RES 100 ML Non-Sterile W CLAMP, TUBING FLAG, & FEMALE LUER, Product Code 68-2230-28
Reason: Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
|
2024-06-26 |
Open, Classified
|
View Details
|
| Z-0018-2025 |
ENT RES, 100 ML, Non-Sterile W CLAMP & FEMALE LUER, Product Code 68-2231-28
Reason: Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
|
2024-06-26 |
Open, Classified
|
View Details
|