Medical Device Recall: Z-0011-2007

FDA Recalls dataset · Firm: Smith & Nephew Inc · Status: Terminated

Recall Action Class: N/A Date Initiated: 2006-08-29 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0011-2007.

Recall Res Number
Z-0011-2007
Recalling Firm / Manufacturer
FEI Number
1020279
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2006-08-29
Date Posted
2006-10-03
Date Terminated
2010-04-12
Product Code
LXH
Event Number
36293
Product Quantity
62 units
Product Description Genesis II Oversized Nonporous Fin-Stem Tibial Punch, Catalog Numbers: 71927102, 71927103, 71927104, 71927105

Reason for Recall Tibial punches were not manufactured to specification and could break at the tip during use.

Action Details Recall notification was initiated by letter (Fed Ex) to domestic territory manager and sales representative and to international distributors (via email and telephone follow-up) on 08/29/2006 to request the return of any in stock product.

Distribution Pattern Worldwide distribution ----- including USA and countries of Austria, Australia, Belgium, Canada, China, Dubai, France, Germany, Italy, Japan, Malaysia, Mexico, Netherlands, Norway, Portugal, Singapore, South Africa, South Korea, Spain, Turkey, United Kingdom, and Venezuela.

Code Information / Serial Numbers Catalog Number 71927102 (Size 1-2) Lot Numbers 06AM00936, 05KM07017, 04KM18886, 04JM05040, 03FM11332, 03FM11334, 03DM01797, 02FM07372, 02CM04108, 01LM08844, 01AM16024, 00306631  Catalog Number 71927103 (Size 3-4) Lot Numbers: 06AM00937, 04LM11514, 04KM18887, 04KM12225, 04FM08554, 03FM11350, 03FM11352, 03DM01798, 02FM07387, 02CM04110, 01LM08847, 00306632  Catalog Number 71927104 (Size 5-6) Lot Numbers: 04KM15385, 04LM09224, 03FM11353, 03DM01800, 02CM04114, 01LM08850, 00306635  Catalog Number 71927105 (Size 7-8) Lot Numbers: 06AM00940, 05KM07018, 04KM18889, 04LM09225, 04FM08555, 03FM11356, 03FM11355, 03DM01803, 02CM04121, 02CM04118, 01LM08854, 01LM08852, 01HM12256, 01BM14137, 00306638

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.