Medical Device Recall: Z-0009-2025

FDA Recalls dataset · Firm: Becton, Dickinson and Company, BD Bio Sciences · Status: Open, Classified

Recall Action Class: N/A Date Initiated: 2024-08-12 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0009-2025.

Recall Res Number
Z-0009-2025
Recalling Firm / Manufacturer
FEI Number
3027394506
Recall Status
Open, Classified
Recall Classification
N/A
Date Initiated
2024-08-12
Date Posted
2024-10-03
Date Terminated
N/A
Product Code
PER
Event Number
95215
Product Quantity
130
Product Description BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSDuet Sample Preparation System (Base), REF: 663128, 662588, used with software V 1.4.1. Intended use: to prepare human specimens to be analyzed.

Reason for Recall Sample preparation system with software may cause: 1) Multidispense feature enabled with 2-tube assay: Specimen volume dispensed into tube1 at least 3% higher than volume in tube2, or 2) Absolute counts calculated using Trucount Tubes: increased volume in tube1 may result in 20% higher absolute counts in tube 1 than tube 2; could result in additional blood collection, delayed results/patient therapy.

Action Details On 8/12/2024, correction notices were sent to customers who were informed of the following; 1) Cease use of the sample preparation system with software version 1.4.1. until the Multidispense feature has been disabled. Patient samples can be prepared using single dispense on the sample preparation system or manually. 2) It is recommended that laboratories upgrade affected sample preparation systems to version 1.4.2 to correct the issue as soon as it is available. 3) Ensure the contents of this notification are read and understood by those within your organization. 4) Complete and return the response form via email to BDRC38@bd.com If you require further assistance, contact: 1-844-823-5433, Mon-Fri 8:00am and 5:00pm CT or email productcomplaints@bd.com. Recall related questions can be emailed to BDRC38@bd.com

Distribution Pattern US: NJ, UT, WI, TX, ID, AK, MN, MA, MD, TN, MI, CA, VA OUS: Ireland, Austria, Italy, Germany, Belgium, Finland, Canada, France, Spain, New Zealand, United Kingdom, Japan, Australia, Norway, Singapore, Sweden, Turkey, Brazil, Colombia, Netherlands, United Arab Emirates, Saudi Arabia, Mexico, Switzerland, Costa Rica, China, Azerbaijan, Algeria, Jordan, India

Code Information / Serial Numbers REF/UDI-DI/Serial Numbers: 666340/382906663405/9230010101, 9230010102, 9230010103, 9230010104, 9230010105, 9230010107, 9230010108, 9230010109, 9230010112, 9230010113, 9230010114, 9230010115, 9230010116, 9230010118, 9230010119, 9230010120. 666339/9230000102, 9230000103, 9230000104, 9230000105, 9230000106, 9230000112, 9230000113, 9230000114, 9230000118, 9230000120, 9230000121, 9230000123, 9230000124, 9230000126, 9230000128, 9230000130, 9230000132, 9230000134, 9230000135, 9230000138, 9230000144, 9230000145, 9230000131, 9230000152, 9230000149, 9230000030, 9230000101, 9230000108, 9230000109, 9230000110, 9230000111, 9230000115, 9230000116, 9230000117, 9230000119, 9230000122, 9230000125, 9230000127, 9230000129, 9230000136, 9230000139, 9230000141, 9230000142, 9230000143. 663128/382906631282/9220030004, 9220030005, 9220030008, 9220030016, 9220030017, 9220030027, 9220030029, 9220030031, 9220030043, 9220030053, 9220030057. 662588/9220010119, 9220010179, 9220010182, 9220010232, 9220010236, 9220010250, 9220010265, 9220010275, 9220010293, 9220010311, 9220010313, 9220010316, 9220010324, 9220010327, 9220010335, 9220010338, 9220010341, 9220010344, 9220010348, 9220010349, 9220010292, 9220010112, 9220010120, 9220010125, 9220010127, 9220010128, 9220010130, 9220010133, 9220010138, 9220010154, 9220010161, 9220010183, 9220010188, 9220010197, 9220010215, 9220010225, 9220010231, 9220010238, 9220010240, 9220010242, 9220010244, 9220010256, 9220010257, 9220010260, 9220010263, 9220010268, 9220010269, 9220010279, 9220010280, 9220010289, 9220010290, 9220010294, 9220010296, 9220010329, 9220010332, 9220010345, 9220010354, 9230000031, 99220010292.

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.