FDA Recall History (1)
| Recall # |
Product Description |
Initiation Date |
Status |
Action |
| Z-0009-2025 |
BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSDuet Sample Preparation System (Base), REF: 663128, 662588, used with software V 1.4.1.
Intended use: to prepare human specimens to be analyzed.
Reason: Sample preparation system with software may cause: 1) Multidispense feature enabled with 2-tube assay: Specimen volume dispensed into tube1 at least 3% higher than volume in tube2, or 2) Absolute counts calculated using Trucount Tubes: increased volume in tube1 may result in 20% higher absolute counts in tube 1 than tube 2; could result in additional blood collection, delayed results/patient therapy.
|
2024-08-12 |
Open, Classified
|
View Details
|