Known Identifiers & Locations
FEI Numbers
2182207
Owner Operator Numbers
N/A
Locations Covered
N/A
Registered Establishments (0)
No device registration and listing records found for this entity.
510(k) Premarket Notifications (0)
No 510(k) notifications found for this entity.
FDA Recall History (3)
| Recall # |
Product Description |
Initiation Date |
Status |
Action |
| Z-0022-2007 |
Medtronic SynchroMed EL Programmable Pumps, Models 8627-10, 8627-18, 8627L-10, 8627L-18. The implantable Medtronic SynchroMed EL Programmable Pump is part of the SynchroMed EL Infusion System designed to contain and administer parenteral drugs to a specific site. The implantable components of the SynchroMed EL Infusion System include the pump with or without a side catheter access port, catheters, and catheter accessories.
Reason: The Catheter Access Port (CAP) on SynchroMed EL pumps manufactured between March and July 1999 may detach from the main body of the pump, which can interrupt drug flow to the target site.
|
2006-06-06 |
Terminated
|
View Details
|
| Z-0037-2008 |
Medtronic InterStim II INS Model 3058 with N'Vision InterStim-B Software supplied on the Model 8870 Version MMB_01/NNB_01 N'Vision with a Clinic an Programmer Model 8840. Implantable Neurostimulator for Urinary Control.
Reason: A software issue may cause a lower than expected battery capacity estimate when evaluating an InterStim II Implantable Neurostimulator (model 3058) with an N'Vision Clinician Programmer(Model 8840) before implant .
|
2007-06-25 |
Terminated
|
View Details
|
| Z-0040-2008 |
Medtronic InterStim INS (Model 3023) with N'Vision Inter Stim-B software supplied on the Model 8870 Version MMB_01/NNB_01 with N'Vision Clinician Programmer Model 8840 or InterStim iCon Patient Programmer Model 3037. Implantable Neurostimulator for Urinary Control.
Reason: Software issue may cause false low battery alert to be displayed when interrogating an Interstim Implantable Neurostimulator (Model 3023) with an N'Vision Clinician Programmer (Model 8840), or with an Inter Stim iCon Patient Programmer (Model 3037).
|
2007-06-25 |
Terminated
|
View Details
|
Profile Disclaimer
AvaSpecs compiles public data records from the FDA database. This compiled profile does not indicate endorsement, safety, clearance, approval, or compliance status. AvaSpecs is a private search index and is not affiliated with the FDA or any government agency.