FDA Recalls dataset · Firm: Alcon Research LLC Aspex Facility · Status: Terminated

Recall Action Class: N/A Date Initiated: 2020-04-17 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-2287-2020.

Recall Res Number
Z-2287-2020
Recalling Firm / Manufacturer
FEI Number
1610287
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2020-04-17
Date Posted
N/A
Date Terminated
2021-02-05
Product Code
HQL
Event Number
85563
Product Quantity
84 units
Product Description Acrysof, IQ IOL w/UltraSert System, AcrySof IQ ASPHERIC IOL, 20.5 D, SP ACRYLIC FOLDABLE LENS, w/UltraSert DELIVERY SYSTEM, UV w/BLUE LIGHT FILTER. 13.0mm. LENGTH, 6.0mm ANTERIOR ASYMMETRIC BICONVEX OPTIC, PLANAR HAPTICS.

Reason for Recall Incorrect IOL diopter

Action Details On 04/17/2020, the firm sent an "Urgent: Voluntary Medical Device Removal' letter to customers via overnight mail that the firm became aware that there was a potential problem relating to the IOL of wrong sized diopter, in a pre-loaded delivery system. Should a patient be implanted with an IOL having incorrect optical power, a refractive error would result. Patients with a refractive error may experience symptoms from mild blurred vision to decreased vision, which may require additional interventions such as spectacles or surgical intervention to correct. The firm is instructing customers to: (i) Return any unused product of specified lot number (ii) The firm's Customer Service will contact customers to arrange for the return and replacement of affected Pre-loaded delivery system (iii) Fill out "Response Form" (iv) Return the "Response form" via fax or email to the firm, provided of contact information: Fax No.: 817-302-4337; E-mail: Market.Actions@alcon.com Additionally, the firm is asking customers to: Reach out to firm's Sales Representative for any questions.

Distribution Pattern US Nationwide distribution including in the states of TX, PA, CT, IA, MI, NE.

Code Information / Serial Numbers Lot number 12726594; Model Number: AU00T0V205; UDI # 038065GMN000065H7; 0380652394772(17)220630(21)12726594000(30)1

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.