Medical Device Recall: Z-2028-2024

FDA Recalls dataset · Firm: Remote Diagnostic Technologies Ltd. Farnborough Aerospace Centre Aerospace Boulevard Farnborough United Kingdom · Status: Open, Classified

Recall Action Class: N/A Date Initiated: 2024-05-09 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-2028-2024.

Recall Res Number
Z-2028-2024
FEI Number
3003832357
Recall Status
Open, Classified
Recall Classification
N/A
Date Initiated
2024-05-09
Date Posted
2024-06-06
Date Terminated
N/A
Product Code
LDD
Event Number
94618
Product Quantity
1745
Product Description Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020 when using Tempus LS-Manual User/Operator Manual

Reason for Recall Simultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference, which may lead to distortions of the ECG signal; distortions may lead to no or ineffective pacing, or no or inappropriate defibrillation pulse.

Action Details On 5/9/24, Philips distributed correction notices to customers and distributors asking them to do the following: 1) The User/Operator Manual was updated can be downloaded by searching "Tempus LS-Manual User-Operator Manual" at Philips.com/IFU 2) This notice needs to be shared with those who need to be aware within your organization and any organization where the potentially affected devices have been transferred. 3) Complete and return the response form via email to ECR.Recall.Response@Philips.com. The following URL can also be used to complete/return the response form: https://philips.efrnfeedback.com/se/705E3ED80E91C689

Distribution Pattern US Nationwide distribution including in the states of VA, DC, MA, WY, AZ, FL, CA, NM, NJ, TN, OH, KS, IN, NY, IL, TX, NV, NE, PA, AL, IA, GA, MO, KY, WI, OR, CO, WV, WA, ET, UT, ID, OK, MN, ME, MT, MI, NC, ND, MD, DE, VT.

Code Information / Serial Numbers 00-3020: UDI-DI-07613365002737, Serial Numbers prior to 7022.002871 User/Operator Manual Versions: 41-3007EN-01 and prior

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.

Account Required Full available specs will require a free AvaSpecs account.