Medical Device Recall: Z-1898-2024

FDA Recalls dataset · Firm: FUJIFILM Healthcare Americas Corporation · Status: Open, Classified

Recall Action Class: N/A Date Initiated: 2024-04-11 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-1898-2024.

Recall Res Number
Z-1898-2024
Recalling Firm / Manufacturer
FEI Number
1000513161
Recall Status
Open, Classified
Recall Classification
N/A
Date Initiated
2024-04-11
Date Posted
2024-05-24
Date Terminated
N/A
Product Code
KPR
Event Number
94443
Product Quantity
18 units
Product Description FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. Model/Catalog Number: 566-16130-23 566-16130-33

Reason for Recall Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury

Action Details FUJIFILM Healthcare Americas Corporation (FUJIFILM) issued Urgent -FIrst Notice letter via email on 4/11/24. Letter states reason for recall, health risk and action to take: 1)Please complete and return the form attached on page 3 of this communication. Providingthe response with the information requested is essential for ensuring appropriate action istaken 2)Please contact Customer Service at 1-888-FUJIMED (1-888-385-4633) to open a serviceticket. 3)Until your upgrade is scheduled and completed, you can continue to safely use FDRVisionary Suite, however, inspect the bolts prior to use. Please contact our field action coordinator, Chaitrali Kulkarni, via email at hcusregulatoryaffairs@fujifilm.com, if you have any further questions regarding this field action.

Distribution Pattern US Nationwide distribution in the states of AK, CA, FL, IL, NE, NH, OH, TN, WI.

Code Information / Serial Numbers UDI-DI: 04540217052226 04540217057450 S/N: MP95A9F6A001, MP95A9F74002, MP95A9F74001, MP95AA983002, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA993001, MP95AA996002, MP95AA996001, MP95AA997001, MP95A9F6B001, MP95A9F76001, MP95AA983001, MP95AA988002, MP95AA98A001, MP95AA993002

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.