Medical Device Recall: Z-1370-2024

FDA Recalls dataset · Firm: Alcon Research, LLC · Status: Open, Classified

Recall Action Class: N/A Date Initiated: 2024-02-21 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-1370-2024.

Recall Res Number
Z-1370-2024
Recalling Firm / Manufacturer
FEI Number
1000117501
Recall Status
Open, Classified
Recall Classification
N/A
Date Initiated
2024-02-21
Date Posted
2024-03-22
Date Terminated
N/A
Product Code
LRO
Event Number
94094
Product Quantity
886
Product Description SYRINGE,BULB,2OZ/60CC,GREEN, Part: 300055070, packaged inside Alcon Custom-Pak Surgical Procedure Packs, Products: 18018-06, 3865-44, 5675-76, 13554-11, 18018-06, 15190-16 , 15190-16, 15190-16, 15190-16, 15190-16, 13554-11, 5918-39, 15190-16, 15190-16, 9208-66, 9208-66, 5675-76, 5918-39, 9208-66, 15190-16, 9065-25, 5918-39, 16066-15, 16066-15

Reason for Recall Medline Bulb irrigation syringes recalled due to potential for packaging sterile barrier to be breached, and these syringes were packed inside surgical procedure packs that were subsequently terminally sterilized, so after the packs are opened, syringes should be discarded at surgical case end, and not reshelved, because the use of non-sterile bulb irrigation syringes may increase infection risk.

Action Details On 2/21/24, correction notices were mailed to customers who were asked to do the following: 1) Review your inventory to determine if you have any affected product within your facility. 2) If you have any affected Custom Pak lots in inventory, please use the included stickers to label any affected product. If you have previously reshelved either of the affected Medline Bulb Irrigation Syringe lots, dispose them. 3) All components inside the Pak are intended to be used only for one specific surgical case and should be discarded at the end of the surgical case. 4) Forward the correction notice to all departments within your organization who may be in possession of affected product; and any other organization to which this product may have been transferred. 5) Complete and return the response form via email to Market.Actions@Alcon.com Customers with questions or concerns can call the firm at 1-800-862-5266.

Distribution Pattern US Nationwide distribution in the states of OH, WV.

Code Information / Serial Numbers Product/Lot/Expiration/UDI-DI: 18018-06/15T707/02/28/2025/380655148440, 3865-44/15T2RC/02/28/2025/380655148150, 5675-76/15T70E/02/28/2025/380655148174, 13554-11/15V4RC/02/28/2025/380655143711, 18018-06/15V3E0/02/28/2025/380655148440, 15190-16/JZ5179517/06/30/2024/380655148303, 15190-16/JZ5179519/06/30/2024/380655148303, 15190-16/JZ5179521/06/30/2024/380655148303, 15190-16/JZ5179523/06/30/2024/380655148303, 15190-16/JZ5179529/06/30/2024/380655148303, 13554-11/15TAUD/01/31/2026/380655143711, 5918-39/15TE5F/02/28/2025/380655149140, 15190-16/15THVA/04/30/2025/380655148303, 15190-16/15TNXJ/04/30/2025/380655148303, 9208-66/15V8CU/04/30/2025/380651693326, 9208-66/15XVDD/05/31/2025/380651693326, 5675-76/JZ5180811/06/30/2024/380655148174, 5918-39/JZ5180812/06/30/2024/380655149140, 9208-66/JZ5180332/03/31/2025/380651693326, 15190-16/JZ5180856/05/31/2025/380655148303, 9065-25/JZ5187593/11/30/2025/380651682849, 5918-39/JZ5187260/06/30/2024/380655149140, 16066-15/JZ5190913/07/31/2025/380655149683, 16066-15/JZ5191891/07/31/2025/380655149683 Bulb Syringe lots: 96922110001 or 96923100001

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.

Account Required Full available specs will require a free AvaSpecs account.