Medical Device Recall: Z-1293-2025

FDA Recalls dataset · Firm: Alcon Research LLC · Status: Open, Classified

Recall Action Class: N/A Date Initiated: 2025-02-10 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-1293-2025.

Recall Res Number
Z-1293-2025
Recalling Firm / Manufacturer
FEI Number
1610287
Recall Status
Open, Classified
Recall Classification
N/A
Date Initiated
2025-02-10
Date Posted
2025-03-06
Date Terminated
N/A
Product Code
LRO
Event Number
96285
Product Quantity
2511
Product Description Alcon Custom Pak with Non-Latex Symbol, Pak: 12771-18, 18050-10, 18252-17, 12698-09, 15632-10, 3141-45, 12630-07, 18925-05, 19265-05, 13894-16, 17550-08, 12236-18, 11854-15, 8043-29, 12466-17, 15599-17, 14764-09, 13194-16, 17069-09

Reason for Recall Customer-designed sterile surgical procedure packs contain latex components as specified; but, the content label sealed within the transparent sterile pouch contains the symbol for a latex-free product. If users or patients with latex sensitivity come into contact with latex, an allergic reaction may occur.

Action Details On 2/10/2025, correction notices were mailed to customers who were asked to do the following: 1) Do not open packaging. Using the provided stickers, adhere one sticker to the outside of affected product near the label to identify it as containing latex. 2) Complete and return the response form electronically via: https://qrco.de/non-latex-symbol Should you have any questions or concerns about this matter, please feel free to call the firm at A 1-800-862-5266

Distribution Pattern US Nationwide distribution in the states of AK, AR, AZ, FL, MD, MN, NC, NE, OH, PA, TX, WA, WI.

Code Information / Serial Numbers UDI-DI: H5301ALCON1CPAK10. Pak/Lot(Expiration): 12771-18/JZ5190507(3/31/2025); 18050-10/JZ5200922(5/31/2025), JZ5202199(9/30/2025); 18252-17/JZ5200852(5/31/2025); 12698-09/JZ5201264(6/30/2025); 15632-10, JZ5202447(6/30/2025), JZ5202445(6/30/2025); 3141-45/ JZ5201303(7/31/2025), JZ5201305(7/31/2025); 12630-07/ JZ5201295(8/31/2025); 18925-05/JZ5194816(9/30/2025), JZ5194818(9/30/2025), JZ5196180(9/30/2025), JZ5196182(9/30/2025), JZ5200323(11/30/2025); 19265-05/JZ5201146(9/30/2025), JZ5201148(9/30/2025), JZ5202207(9/30/2025); 13894-16/JZ5201181(9/30/2025), JZ5202143(9/30/2025); 17550-08/JZ5198232(10/31/2025); 12236-18/JZ5201054(11/30/2025); 11854-15/JZ5201561(11/30/2025); 8043-29/JZ5202201(11/30/2025); 12466-17/JZ5205311(12/31/2025), JZ5205313(12/31/2025), JZ5205915(1/31/2026); 15599-17/JZ5205311(12/31/2025), JZ5205313(12/31/2025), JZ5205915(1/31/2026); 14764-09/JZ5197158(2/28/2026); 13194-16/JZ5197142(3/31/2026); 17069-09/JZ5201931(9/30/2026)

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.