Medical Device Recall: Z-0895-2018

FDA Recalls dataset · Firm: Instrumentation Laboratory Co. · Status: Terminated

Recall Action Class: N/A Date Initiated: 2017-08-02 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0895-2018.

Recall Res Number
Z-0895-2018
Recalling Firm / Manufacturer
FEI Number
1217183
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2017-08-02
Date Posted
2018-03-03
Date Terminated
2019-03-29
Product Code
GJS
Event Number
79165
Product Quantity
14,894 kits
Product Description Hemosll ReadiPlasTin (20 ml Size), Part Number 0020301400

Reason for Recall Instrumentation Laboratory Co. received customer reports of performance issues with some vials of Lot Nos. N1166235 and N0177760, including increased imprecision, out of range quality controls and prolonged sample results.

Action Details On August 2, 2017, Instrumentation Laboratory sent an Urgent Medical Device Removal letters dated August 2, 2017 to their customers. On October 5, 2017 Instrumentation Laboratory expanded their recall adding 3 additional lots to a second Urgent Medical Device Removal letter. On February 26, 2018, Instrumentation Laboratory expanded their recall a 3rd time to include 1 additional lot to a third Urgent Medical Device Removal letter. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete a Mandatory Tracking Response form. For questions contact your Customer Service Representative at 1-800-955-9525, option #1.

Distribution Pattern Worldwide Distribution - US Nationwide

Code Information / Serial Numbers Lots: N1166235, N0177760, N0278358, N0478057, N0479504 & N0278177.

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.