Medical Device Recall: Z-0605-2010

FDA Recalls dataset · Firm: Abbott Medical Optics Inc (AMO) · Status: Terminated

Recall Action Class: N/A Date Initiated: 2009-11-11 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0605-2010.

Recall Res Number
Z-0605-2010
Recalling Firm / Manufacturer
FEI Number
3003843509
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2009-11-11
Date Posted
2010-01-12
Date Terminated
2012-07-25
Product Code
HQL
Event Number
53781
Product Quantity
246000 Lenses
Product Description AMO Tecnis 1-Piece Intraocular Lens (Model ZCB00)

Reason for Recall Abbott Medical Optics (AMO) has received complaints of some Tecnis 1-Piece IOLs (model ZCB00) have adhered to the lens optic after insertion into the eye. This can lead to a need for additional manipulation to separate the haptics from the optic during surgery which could lead to an injury.

Action Details Beginning on Friday, November 6, 2009, AMO began distribution of the Recall Notification Letter via Federal Express to the 874 AMO Tecnis 1-piece IOL customer accounts in the US. Included with the Recall Notification letter is a facsimile delivery confirmation that each customer was instructed to fax back to AMO to confirm receipt of the Recall Notification Letter. --- Beginning on Monday, November 9, 2009, AMO began distribution of the Recall Notification letter to the AMO Tecnis 1-piece IOL customer/distributors accounts outside the USA. "THIS RECALL NOTIFICATION DOES NOT INVOLVE TECNIS 1-PIECE IOL THAT HAVE BEEN IMPLANTED. ONCE THE HAPTICS HAVE BEEN SUCCESSFULLY DEPLOYED, THESE LENSES FUNCTION PROPERLY." Questions should be directed to an AMO Customer Service Representative at 1-877-AMO-4LIFE (1-877-266-4543).

Distribution Pattern AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY, Hawaii, and Puerto Rico. INTERNATIONAL: Austria, Azerbaijan, Belgium, Canada, Switzerland, Cyprus, Czech Republic, Germany, Estonia, Spain, Finland, France, United Kingdom, Hungary, Ireland, Israel, Italy, Lebanon, Lichtenstein, Libya, Martinique, Netherlands, Poland, Palestine, Portugal, Saudi Arabic, Sweden, Slovenia, Turkey, and South Africa.

Code Information / Serial Numbers All Serial numbers of tecnis 1-Piece Intraocular Lenses that were processed through lens surface treatment at the AMO PR facility between May 19 and October 16, 2009 - with serial numbers ending in 0906, 0907, 0908, and 0909. and subset with serial numbers ending in 0905 and 0910.  Expiration/Use By Date (YYYY-MM): 2011-05, 2011-06, 2011-07, 2011-08, 2011-09, and 2011-10

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.