Medical Device Recall: Z-0471-2022

FDA Recalls dataset · Firm: Philips North America Llc · Status: Terminated

Recall Action Class: N/A Date Initiated: 2021-11-12 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0471-2022.

Recall Res Number
Z-0471-2022
Recalling Firm / Manufacturer
FEI Number
3016618143
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2021-11-12
Date Posted
N/A
Date Terminated
2024-02-05
Product Code
OWB
Event Number
89174
Product Quantity
10 units (1 US and 9 OUS)
Product Description Philips Azurion Interventional Fluoroscopic X-ray System with Azurion R2.2 with LUC configuration start-up issue; Azurion R2.2 emergency stop recovery issue Model Numbers: 722221, 722228, 722225, 722068, 722226, 722078, 722224

Reason for Recall (1)Start-up problem: Intermittently at start-up of the system, the communication of the control unit that manages the movement of the stand is not established; (2) Emergency stop recovery problem-warm restart is not executed correctly and movement functionality is not available

Action Details Philips issued Field Safety Notice on 11/12/21 U.S.A, Customer notified via certified mailing from the United States Postal Service (USPS). Letters outside of the U.S.A. will be distributed through the Philips Markets Organization. Letter states reason for recall, health risk and action to take: Start-up problem: If the start-up problem occurs, perform a Cold Restart* of the system. b) Emergency stop recovery problem: If the emergency stop recovery problem occurs, perform a new Warm Restart* of the system. c) Place this Field Safety Notice with the documentation of the system until Philips has installed a software update in your system. d) Circulate this notice to all users of this device so they are aware of the product issues. e) Return the attached reply form to Philips to confirm that the users of the system have reviewed and understood this Field Safety Notice. * The system restart options are described in section 4.2 of the Instructions for Use. 5. Describe the actions planned by Philips IGT Systems to correct these problems: These problems will be resolved by a software update (available in November 2021). You will be contacted by your local Philips representative to schedule the update of your system. If you need any further information, please contact your local Philips representative (reference to FC072200502).

Distribution Pattern Worldwide distribution - US Nationwide distribution in the state of WA and the countries of France, Germany, Japan, Poland, United Kingdom.

Code Information / Serial Numbers US: Model 722068 Azurion 7 B20 S/N: 109 UDI: (01)00884838085367(21)109. OUS: Model: S/N UDI: 722226 Azurion 7 B20 176 (01)00884838099272(21)176; 722228 Azurion 5 M20 3 (01)00884838099234(21)3; 722221 Azurion 3 M12 8 (01)00884838099203(21)8; 722225 Azurion 7 B12 16 (01)00884838099265(21)16; 722225 Azurion 7 B12 73 (01)00884838099265(21)73; 722226 Azurion 7 B20 42 (01)00884838099272(21)42; 722226 Azurion 7 B20 157 (01)00884838099272(21)157; 722224 Azurion 7 M20 293 (01)00884838099258(21)293; 722078 Azurion 7 M12 220 (01)00884838085251(21)220

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.