Medical Device Recall: Z-0462-2023

FDA Recalls dataset · Firm: Philips North America Llc · Status: Open, Classified

Recall Action Class: N/A Date Initiated: 2022-11-22 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0462-2023.

Recall Res Number
Z-0462-2023
Recalling Firm / Manufacturer
FEI Number
3016618143
Recall Status
Open, Classified
Recall Classification
N/A
Date Initiated
2022-11-22
Date Posted
2022-12-08
Date Terminated
N/A
Product Code
LNH
Event Number
91195
Product Quantity
566 total
Product Description Philips MR 7700 System, Model Number 782120

Reason for Recall Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.

Action Details An URGENT Medical Device Correction Notice dated 11/22/22 was sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users 1. Systems that do not yet have the SmokeDetector Interlock feature installed are identified in Appendix A. If your system is listed in Appendix A immediately discontinue use of your impacted MR system(s) until Philips has installed the SmokeDetector Interlock. " Post a Do not use notice on or near the impacted MR system(s) " Post this notice near the affected MR system(s) for ease of reference. " Circulate this notice to all users of this device so that they are aware of the product issue and associated hazard/harm until this issue has been resolved. " Please complete and return the attached customer response form to Philips promptly and no later than 30 days from receipt of this letter. " Note: Once your system has the SmokeDetector Interlock installed, please follow the advice in steps 2 through 5. 2. Philips has installed a SmokeDetector Interlock feature on the systems listed in Appendix B. Immediately confirm if your system has the SmokeDetector Interlock installed by confirming your serial number is listed in Appendix B. " Post this notice near the affected MR system(s) for ease of reference. " Circulate this notice to all users of this device so that they are aware of the product issue and associated hazard/harm until this issue has been resolved. " Please complete and return the attached customer response form to Philips promptly and no later than 30 days from receipt of this letter. 3. The SmokeDetector Interlock feature is integrated with the MR System and powers off the relevant part of the system (Gradient Amplifier) when smoke is detected in order to prevent further development of smoke and/or fire. If the SmokeDetector Interlock has detected smoke and has powered off the Gradient Amplifier: " Remove patient from the system according to the instructions for use because

Distribution Pattern US Nationwide.

Code Information / Serial Numbers Model 782120: UDI-DI: 00884838104112; Serial Numbers: 65005 65006 65007 65008

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.