Medical Device Recall: Z-0355-2022

FDA Recalls dataset · Firm: Philips North America Llc · Status: Open, Classified

Recall Action Class: N/A Date Initiated: 2021-11-12 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0355-2022.

Recall Res Number
Z-0355-2022
Recalling Firm / Manufacturer
FEI Number
3016618143
Recall Status
Open, Classified
Recall Classification
N/A
Date Initiated
2021-11-12
Date Posted
N/A
Date Terminated
N/A
Product Code
OWB
Event Number
89129
Product Quantity
22 units US
Product Description BV Endura with Software Release 2.3- A mobile, diagnostic X-ray image acquisition and viewing system. Model Number: 718075

Reason for Recall Instructions for Use do not specify the maximum surface temperature of the X-ray tank (assembly housing the X-ray tube, beam filter & cooling oil) and the image intensifier/detector of the system as required an result in burns

Action Details Philips issued an Urgent Medical Device Correction letter in the USA on 11/12/21 via Certified mailing United States Postal Service. Recalling outside the US is done through the local Philips organizations in the respective countries Letter states reason for recall, health risk and action to take: Pay attention to the Heat Indication Level Icons shown in the Viewing Station to know the temperature of the X-Ray Tank. o Follow the instructions provided in Annex I of this letter for a safe use of the system. o Circulate this notice to all users of this device so that they are aware of the issue. o Place this Field safety Notice with the documentation of the of your system o Transfer this Field Safety notice within your organization or to any organization, if the affected devices have been transferred. o Complete the enclosed Customer Reply Form and send it back to confirm that users have reviewed and understood this Urgent Field Safety Notice. 5. Actions planned by Philips IGT Systems to correct the problem Philips is informing customers through this Urgent Field Safety Notice and providing the Addendum to the Instructions For Use of the BV Endura Release 2.3, BV Pulsera Release 2.3 and Veradius Unity Release 2.1 systems.

Distribution Pattern Worldwide distribution - US Nationwide and the countries of Afghanistan, Algeria, American Samoa, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Cameroon, Chile, China, Colombia, Cook Islands, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Faroe Islands, Finland, France, French Guiana, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guatemala, Guyana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kuwait, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Macedonia, Malaysia, Maldives Mali, Martinique, Mauritius, Mexico, Moldova, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of Panama, Papua New Guinea, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia, Zimbabwe.

Code Information / Serial Numbers Serial Number: 173 105 5301 4950 5360 4651 4891 873 1633 160 543 562 204 5090 5091 5092 4840 5073 125 383 384 5121 UDI: (01)00884838055490(21)173 (01)00884838055490(21)105 (01)00884838055490(21)5301 (01)00884838055490(21)4950 (01)00884838055490(21)5360 (01)00884838055490(21)4651 (01)00884838055490(21)4891 (01)00884838055490(21)873 (01)00884838055490(21)1633 (01)00884838055490(21)160 (01)00884838055490(21)543 (01)00884838055490(21)562 (01)00884838055490(21)204 (01)00884838055490(21)5090 (01)00884838055490(21)5091 (01)00884838055490(21)5092 (01)00884838055490(21)4840 (01)00884838055490(21)5073 (01)00884838055490(21)125 (01)00884838055490(21)383 (01)00884838055490(21)384 (01)00884838055490(21)5121

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.