Product Description
Covidien Sonicision Battery Charger-intended to charge the Sonicision Reusable Battery Pack and to provide information on the battery charging progress and battery pack status
Item Code: CBCA
Reason for Recall
Intermittent risk of electric shock to users if an affected charger is used in conjunction with a faulty or bypassed institutions Mains Protective Earth (Ground) circuit
Action Details
Medtronic customer communication translated by in-country representatives emailed on November 12, 2021. Letter states reason for recall, health risk and action to take:
1. Please immediately quarantine and discontinue use of the affected item code with the associated serial numbers listed above.
2. Please return affected product as indicated below. All products from the affected item code and associated serial numbers must be returned.
3. If you have distributed the Covidien Sonicision" Battery Chargers listed above, please promptly forward the information from this letter to those recipients.
4. Complete the Recalled Product Return Form even if you do not have inventory.
Fax to 800-895-6140 or email to rs.covidienfeedbackcustomerservice@medtronic.com
Credit for returned affected product will be issued based on the RGA number.
Exception: Customers with zero inventory, fax to 651-367-2624 or email to rs.gmbfcamitg@medtronic.com
Distribution Pattern
Brazil
Code Information / Serial Numbers
Serial Numbers: MCSB006939, MCSB006947
GTIN: 10884521592896
Public Record Disclaimer
This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.