Medical Device Recall: Z-0148-2015

FDA Recalls dataset · Firm: Siemens Healthcare Diagnostics, Inc. · Status: Terminated

Recall Action Class: N/A Date Initiated: 2014-09-24 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0148-2015.

Recall Res Number
Z-0148-2015
Recalling Firm / Manufacturer
FEI Number
2517506
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2014-09-24
Date Posted
2014-10-29
Date Terminated
2016-08-17
Product Code
DIP
Event Number
69372
Product Quantity
1733
Product Description siemens Dimension Phenytoin Flex Reagent Cartridge The PTN method used on the Dimension clinical chemistry system is an in vitro diagnostic test intended to measure phenytoin (dilantin, diphenylhydantoin), an anti-epileptic drug, in human serum and plasma. PTN test results are used in the diagnosis and treatment of phenytoin overdose and in monitoring levels of phenytoin to ensure appropriate therapy.

Reason for Recall Internal testing has confirmed a low absorbance range between the levels 1 and 5 calibrators with this lot resulting in the potential for a positive or negative bias and imprecision across the PTN assay range.

Action Details Siemens sent an Urgent Medical Device Recall Notification dated September 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed as follows: " Discontinue use and discard any remaining inventory of PTN lot FA5154, and complete the attached form indicating your replacement product needs in the attached Field Correction Effectiveness Form. " Siemens will replace any unused inventory of the affected lot at no charge. We apologize for the inconvenience this situation may have cause. If you have any questions, please contact Siemens Customer Care Center - Technical Solutions at (800) 441-9250, or your local Siemens technical support representative.

Distribution Pattern Worldwide Distribution - USA (nationwide) including Puerto Rico., and the countries of Australia, Brazil, Bahamas, Canada, Chile, India, Japan and Uruguay

Code Information / Serial Numbers lot FA5154, exp. 2015-06-03

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.