Medical Device Recall: Z-0070-2019

FDA Recalls dataset · Firm: COVIDIEN LLC · Status: Terminated

Recall Action Class: N/A Date Initiated: 2018-08-17 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0070-2019.

Recall Res Number
Z-0070-2019
Recalling Firm / Manufacturer
FEI Number
1282497
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2018-08-17
Date Posted
N/A
Date Terminated
2020-04-10
Product Code
GAG
Event Number
81043
Product Quantity
1,902
Product Description EEA Circular Stapler with Tri-Staple Technology, Part No. TRIEEA28MT TRIEEA28XT TRIEEA31MT TRIEEA31XT Product Usage: The EEA Circular Stapler with Tri-Staple Technology places a circular, triple staggered row of titanium staples and resects the excess tissue, creating a circular anastomosis.

Reason for Recall The firm identified the potential for a device to have an incorrect tissue gap. Use of a device with an incorrect tissue gap may result in incomplete staple formation and/or the inability to remove the device from tissue following application potentially leading to bleeding, anastomotic leak or tissue trauma. There have been no reports of serious injury related to this issue.

Action Details All U.S. consignees were notified via Federal Express or certified mail on August 17, 2018. The letter informs customers of the potential for use of a device to have an incorrect tissue gap. Required Actions: 1. Please quarantine and discontinue use of the affected item code and lots listed above. 2. Please return affected product. All unused products from the affected item codes and lots must be returned. If you purchased directly from Medtronic, return all inventory of the affected lots to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473, Attention: Field Returns Department. If you purchased from a distributor, complete all fields on the form and contact your distributor directly to arrange for return of product. 3. If you have distributed EEA" staplers with Tri-Staple" technology listed above, please promptly forward the information from this letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.

Distribution Pattern Worldwide Distribution - US Nationwide in the states of CA, FL, IL, NC, NY, UT. and the countries of Austria, Belgium, Denmark, Finland, France, Germany, Iceland, Italy, Latvia, Netherlands, Norway, Poland, Portugal, Reunion, Spain, Sweden, Switzerland, and United Kingdom.

Code Information / Serial Numbers P8A0976X P8A0977X P8A0978X P8A0979X P8A0980X P8A0981X P8B1538X P8B1539X P8B1540X P8B1541X P8B1542X P8C0032X P8C0033X P8C1177X P8C1178X P8C1179X P8C1180X P8C1181X P8C1183X P8D1405X P8D1406X P8D1407X P8D1408X P8D1409X P8D1410X P8D1411X P8E1211X P8E1212X P8E1213X P8E1214X P8E1215X P8E1216X P8E1217X P8E1304X P8E1305X

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.