Medical Device Recall: Z-0044-2026

FDA Recalls dataset · Firm: FUJIFILM Healthcare Americas Corporation · Status: Open, Classified

Recall Action Class: N/A Date Initiated: 2025-09-15 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0044-2026.

Recall Res Number
Z-0044-2026
Recalling Firm / Manufacturer
FEI Number
1000513161
Recall Status
Open, Classified
Recall Classification
N/A
Date Initiated
2025-09-15
Date Posted
2025-10-07
Date Terminated
N/A
Product Code
KPR
Event Number
97581
Product Quantity
16 units
Product Description FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33

Reason for Recall It was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not work effectively only when the shaft was broken while being operated toward the front while lifting.

Action Details On September, 15, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: Perform the daily inspection described in the operation manual to check for wobbles or shakes, misalignment of the light irradiation position, etc. If any abnormality is found, discontinue use and contact Technical Assistance Center (TAC). When rotating the X-ray tube unit, do not hit the stopper at the end. Do not rotate the X-ray tube over or near the patient. Please post this notice on or near the device. Complete and return the response form provided. Fujifilm Service Engineers will install countermeasure parts that will prevent the X-ray tube support from coming off even if the shaft is damaged. Please contact Technical Assistance Center (TAC) at 1-888-385-4633 or 1-888-FUJIMED for any functional/correction questions as well as to schedule your correction.

Distribution Pattern Domestic: AK, CA, FL, IL, MA, NH, OH, WI.

Code Information / Serial Numbers Model/Version Number: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33; UDI-DI: (1) 04540217052226, (2) 04540217057436, (3) 04540217057450; Serial Numbers: (1) MP95A8F5A001, MP95A9F6A001, MP95A9F6B001, MP95A9F74001, MP95A9F74002, MP95A9F76001, MP95A9F78001, (2) MP95AA783001, (3) MP95AA983001, MP95AA983002, MP95AA983003, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA988002;

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.