Product Description
AssayTip/AssayCup Tray-used on the cobas e 801 module which is a part of the cobas c 8000 MODULAR Analyzer Series (IVD for in vitro determination of analytes in human body fluids)
Catalog Number: 5694302001
Reason for Recall
AssayTips part of the AssayTip/AssayCup Tray Tips
used on the cobas e 801 module- may have an Abnormal Internal Structure and lead to Incorrect patient results
Action Details
Roche Diagnostics issued Urgent Medical Notification letter to Distributor (TP-00744) and end users( TP-00741) on 8/13/19. Letter identifies problem, health risk and Actions required per the UMDC:
" Check to see if you have any of the affected lot number(s) of
AssayTips/AssayCup Trays on your cobas e 801 module or in your laboratory
stock. (See pictures for lot number locations on packaging.).
" Stop using the affected lot number(s) of AssayTips/AssayCup Trays immediately.
" Discard the affected lot number(s) of AssayTips/AssayCup Trays according to your local waste management guidelines.
To receive replacement product, fill out the attached fax form and fax or email it according to the instructions on the form.
Contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions
Distribution Pattern
Nationwide
Code Information / Serial Numbers
Lot Numbers: 18636170 18665170 18667170 18668170 18215274 18245274 18250274 18251274 18252274 18253274 18265274 18271274 19209274
Public Record Disclaimer
This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.