Medical Device Recall: Z-0041-2025

FDA Recalls dataset · Firm: Inspire Medical Systems Inc. · Status: Open, Classified

Recall Action Class: N/A Date Initiated: 2024-07-31 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0041-2025.

Recall Res Number
Z-0041-2025
Recalling Firm / Manufacturer
FEI Number
3007666314
Recall Status
Open, Classified
Recall Classification
N/A
Date Initiated
2024-07-31
Date Posted
2024-10-09
Date Terminated
N/A
Product Code
MNQ
Event Number
95230
Product Quantity
457 units
Product Description Inspire Model 4063 Stimulation Lead, component of the Inspire Upper Airway Stimulation (UAS) system.

Reason for Recall Labeling error with a mismatch between the shelf box serial number label and the serial number of the lead provided in the box.

Action Details Inspire notified consignees on 07/31/2024, via emailed letter. Consignees were instructed to confirm the serial number on the lead is the serial number used to register the patient and to contact your Inspire representative if any labeling inconsistency is identified for complaint reporting.

Distribution Pattern Worldwide distribution - US Nationwide and the countries of Belgium, Switzerland, Germany, United Kingdom, Netherlands, Singapore.

Code Information / Serial Numbers Model 4063. UDI-DI: 10855728005793. Part Number: 900-013-001. Serial Numbers: T52648, T52644, T52642, T52640, T52635, T52630, T52634, T52615, T52611, T52654, T52655, T52607, T52604, T52609, T52603, T52595, T52597, T52616, T52594, T52591, T52590, T52592, T52612, T52588, T52625, T52579, T52566, T52561, T52522, T52523, T48445, T48424, T48403, T48388, T48357, T48352, T48345, T48340, T48331, T48332, T48333, T48288, T11952, T11955, T48282, T48265, T48255, T48253, T48254, T48217, T48188, T48184, T48167, T48189, T48213, T48166, T48161, T48153, T48149, T48144, T47959, T47947, T47940, T48435, T47936, T47820, T47818, T47805, T47785, T47772, T47770, T47761, T47748, T47747, T47733, T47731, T47708, T47709, T47683, T47660, T47767, T47768, T47630, T47617, T47615, T47614, T47490, T47473, T47924, T47459, T47447, T47417, T48170, T47400, T47389, T45700, T47824, T47368, T47364, T47354, T43971, T48314, T47349, T47356, T47415, T47342, T47303, T47211, T47440, T46791, T47196, T47186, T47187, T47198, T47200, T43079, T47149, T47144, T47167, T47138, T48262, T47132, T47116, T47206, T47067, T46826, T47674, T46821, T46820, T46802, T46799, T46798, T46797, T46760, T46742, T46714, T46700, T46596, T46592, T46588, T46562, T46563, T46559, T46526, T46353, T46336, T46313, T52538, T46300, T52515, T52520, T46295, T45740, T45736, T45720, T45701, T45697, T45679, T45420, T45405, T45375, T45372, T45366, T45358, T45347, T45333, T45132, T44982, T44965, T44933, T44931, T44911, T44754, T44698, T43947, T43938, T43918, T43910, T43902, T43898, T43881, T43873, T43799, T43795, T43784, T43763, T43754, T43741, T43716, T43732, T43704, T43516, T43506, T43361, T43362, T43363, T43328, T43338, T43326, T43322, T43323, T43101, T43115, T43117, T43299, T43291, T43264, T43249, T43244, T43238, T43236, T43135, T43108, T43096, T30899, T30916, T47710, T47711, T47712, T43093, T43064, T42831, T42814, T42805, T43273, T42794, T42789, T42781, T42760, T42574, T42592, T42593, T42568, T42323, T42122, T47758, T47786, T41973, T52549, T41953, T41954, T41944, T41926, T41927, T41921, T41577, T46721, T46723, T48356, T48360, T10211, T10287, T41920, T41918, T41824, T41780, T43761, T41759, T48215, T41724, T41697, T29941, T29951, T44953, T42298, T42299, T41671, T41649, T47826, T41547, T47312, T41434, T41425, T41670, T41707, T41420, T44681, T47136, T41296, T43889, T43969, T41259, T41277, T41249, T41244, T41195, T42319, T41063, T40951, T41040, T40983, T40959, T40957, T40917, T40849, T40846, T40762, T40763, T40725, T40573, T40708, T40560, T29971, T40110, T40100, T40095, T40043, T38150, T40014, T39976, T39936, T39904, T39889, T39845, T39781, T39827, T39828, T39779, T39769, T39705, T39693, T46365, T39689, T39969, T39670, T37168, T39651, T39625, T39585, T39529, T39518, T39448, T43296, T39443, T39372, T39374, T48363, T48364, T39351, T39326, T42812, T45718, T39313, T39285, T37755, T39933, T39281, T39290, T39267, T39268, T39270, T39271, T39143, T39096, T40101, T39085, T38193, T38203, T39523, T52610, T39081, T39076, T39077, T38984, T42103, T38945, T38450, T38434, T38197, T52651, T52652, T38185, T39530, T38180, T38133, T43455, T38103, T38151, T37879, T37845, T37831, T37830, T37824, T37813, T37800, T36459, T45354, T37780, T40096, T38205, T38206, T37779, T37725, T40124, T37705, T37634, T37589, T37461, T37280, T44750, T37251, T37231, T37216, T43744, T46725, T37200, T37193, T37153, T37058, T36679, T36657, T36614, T36528, T36512, T36513, T36514, T36516, T36517, T36504, T36501, T36458, T36463, T36376, T30914, T30912, T30847, T30851, T30852, T29965, T29964, T29955, T29954, T29953, T29936, T47918, T29935, T27236, T15775, T11827, T11589, T11422, T11297, T11214, T11215, T11614, T11201, T11202, T11213, T11419, T10102, T10131, T10089, T10098, T10099, T10101, T11165, T11166.

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.