Product Description
Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5)
Product Number: ROB10299
Reason for Recall
CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus to 7 degrees varus) when the surgeon chooses to define the femoral rotational reference via the posterior condylar axis
Action Details
Smith & Nephew, Inc. (Blue Belt Technologies) issued Urgent Medical Device Correction Notice (C-2023-08) on September 7, 2023 via overnight mail. Letter states reason for recall, health risk and action to take:
Required Actions for user/hospital:
1. Complete Correction Notice Acknowledgement Form and submit to FieldActions@smith-nephew.com.
2. Prior to use of software, contact your sales representative to update settings.
3. Upon completion of settings update, you may use the CORI Surgical System until the software update is deployed.
4. A Smith & Nephew Robotics Representative will be deployed to your site to implement a new software version to correct the issue once it is available.
Email: FieldActions@smith-nephew.com
Actions for Smith & Nephew Representatives
1. Work with customer to update CORI settings to prevent incorrect configuration.
2. Complete the attached Correction Notice Acknowledgement Form.
3. Upon completion of software upgrade, submit CORI In-Field Service Record to Field Action Team.
Distribution Pattern
US Nationwide distribution.
Code Information / Serial Numbers
UDI: (01) 00885556773338
Lot Numbers: 51141585, 51108475, 51108689
Public Record Disclaimer
This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.