Medical Device Recall: Z-0041-2008

FDA Recalls dataset · Firm: Naviscan PET Systems · Status: Terminated

Recall Action Class: N/A Date Initiated: 2007-08-01 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0041-2008.

Recall Res Number
Z-0041-2008
Recalling Firm / Manufacturer
FEI Number
3005945241
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2007-08-01
Date Posted
2007-10-11
Date Terminated
2008-05-14
Product Code
KPS
Event Number
44791
Product Quantity
19 units
Product Description PEMView Software used with Naviscan PEM Flex Solo I and PEM Flex Solo II PET Scanner, for nuclear medicine imaging.

Reason for Recall Naviscan has identified a potential safety issue with all versions of PEM View Software commonly used for analyzing patient images acquired on the PEMFIex product line. 1. The Positron Emission Mammography Standardized Uptake Value (PEMSUV) and distance measurements for the product mentioned above may be inaccurate if a zoomed image has been opened, measurement tools are used, and the applet rema

Action Details Telephone notification regarding issues #1 and #2 were carried out on August 1 and 2, 2007 and included instructions for a validated work-around to avoid the software issues. Written notifications with these same validated work-around instructions were sent on August 2, 2007 via certified mail. Customers were notified that Naviscan has identified a potential safety issue with all versions of PEM View Software. Issue #3 was identified after the notification to customers for issues #1 and #2. Telephone notification regarding issues #3 was carried out via telephone and certified mailing beginning August 16, 2007. Customers were informed that Naviscan is preparing software to resolve these issues and that they would be contacted by Naviscan Customer Support to ensure they thoroughly understand the issues and the recommended procedure to prevent these issues from occurring. If customers have any questions they were instructed to contact Naviscan Customer Support at 1-888-628-4159 or service@navsicanpet.com.

Distribution Pattern Nationwide.

Code Information / Serial Numbers Serial numbers: C3, P1, P2, P3, P4, P5, P6, P7, P8, P10, P11, P12, P14, P002-B-070119-01, P002-B-070316-01, P002-B-070423-01, P002-B-070531-01, P002-B-070629-01, P002-B-070712-01

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.