Medical Device Recall: Z-0039-2012

FDA Recalls dataset · Firm: Richard-Allan Scientific Company · Status: Terminated

Recall Action Class: N/A Date Initiated: 2011-08-12 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0039-2012.

Recall Res Number
Z-0039-2012
Recalling Firm / Manufacturer
FEI Number
1831638
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2011-08-12
Date Posted
2011-10-12
Date Terminated
2013-07-25
Product Code
IDO
Event Number
59316
Product Quantity
786
Product Description Thermo Scientific Microm: Rotary Microtome: REF HM355S-3 Used for slicing tissue blocks.

Reason for Recall The firm has received two complaints of finger lacerations in operators of this device.

Action Details ThermoFisher Scientific sent "URGENT MEDICAL DEVICE CORRECTION, RE: Thermo Scientific HM355S-3 User Safety Advisory Notice" to all affected customers. The letter states the manufacturer identified a need to enhance the safety instructions and the description of the safety devices in the user manual. It mentions the HM3555S-3 User Manual has been modified to include more specific safe work practices and a new Quick -Reference Card summarizing the manual's safety enhancements and that the changes have been included on a CD that is enclosed with the letter. The firm requests confirmation of receipt of the new User Manual and Quick-Reference Card by October 31, 2011 to be sent by FAX to (269) 372-2674 For questions on this recall call the firms Technical Support representatives at 1-800-522-7270 08:00 AM- 4:30 PM.

Distribution Pattern Worldwide Distribution: Nationwide distribution; including the country of Canada.

Code Information / Serial Numbers Model HM355S-3, Serial Nos: 34450, 34519, 34520, 34601, 34662, 34665, 34807, 34922, 35030, 35032, 35033, 35159, 35637, 35793, 35976, 35978, 35978, 38337, 38584, 38651, 39238, 39280, 39281, 39542, 39795, 40607, 41039, 45104, 45367, 45369, 45371, 45760, 45845, 47466, 47467, 47468, 47469, 47772, 47773, 47774, 47775, 47776, 48095, 48096, 48097, 48098, 48109, 48110, 48111, 48112, 48113, 48114, 48115, 48116, 48184, 48252, 48253, 48254, 48259, 48260, 48261, 48314, 48315, 48593, 48594, 48910, 48911, 48912, 48913, 48914, 48915, 48916, 48917, 48918, 48919, 48920, 48921, 48922, 48923, 48924, 48978, 48978, 49102, 49103, 49104, 49105, 49106, 49107, 49108, 49109, 49110, 49209, 49209, 49210, 49210, 49211, 49211, 49212, 49213, 49213, 49214, 49214, 49221, 49223, 49230, 49286, 49287, 49288, 49289, 49290, 49291, 49292, 49293, 49294, 49295, 49296, 49297, 49298, 49299, 49300, 49301, 49302, 49303, 49304, 49305, 49306, 49307, 49308, 49309, 49310, 49311, 49312, 49962, 49963, 49964, 49965, 49966, 49967, 49968, 49969, 49970, 49971, 49972, 49973, 49974, 49975, 49976, 49977, 49978, 49989, 49990, 49991, 49992, 49993, 49994, 49995, 50221, 50222, 50224, 50225, 50226, 50578, 50579, 50580, 50581, 50582, 50583, 50584, 50585, 50586, 50587, 50676, 50677, 50678, 50679, 50680, 50681, 50791, 50792, 50793, 50794 and 50911.

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.