Medical Device Recall: Z-0038-2010

FDA Recalls dataset · Firm: Airgas Puritan Medical Puritan Medical Unit · Status: Terminated

Recall Action Class: N/A Date Initiated: 2009-08-06 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0038-2010.

Recall Res Number
Z-0038-2010
Recalling Firm / Manufacturer
FEI Number
N/A
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2009-08-06
Date Posted
2009-10-14
Date Terminated
2010-01-26
Product Code
CAN
Event Number
53081
Product Quantity
208
Product Description AGI-2238 Portable 4 Cylinder Manifold Cart, for connecting medical gas cylinders.

Reason for Recall Yoke may become displaced. Yoke index safety system could become compromised, allowing a user to attach the wrong gas type to the yoke connection.

Action Details Airgas issued a Medical Device Removal letter dated 8/6/09 to its customers informing them of the problem and the need to have a yoke back-up plate installed on the Manifold Block Assembly.

Distribution Pattern Nationwide distribution.

Code Information / Serial Numbers All units of model numbers 8065-870 and 8066-910

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.