Medical Device Recall: Z-0032-2026

FDA Recalls dataset · Firm: Beckman Coulter Inc. · Status: Open, Classified

Recall Action Class: N/A Date Initiated: 2025-08-04 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0032-2026.

Recall Res Number
Z-0032-2026
Recalling Firm / Manufacturer
FEI Number
2050012
Recall Status
Open, Classified
Recall Classification
N/A
Date Initiated
2025-08-04
Date Posted
2025-10-02
Date Terminated
N/A
Product Code
CGR
Event Number
97390
Product Quantity
116
Product Description DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i Clinical Analyzer

Reason for Recall On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays.

Action Details On 8/4/2025, recall notices were mailed and/or emailed to customers who were asked to do the following: Check your software version, select Menu - About page. - For Clinical Analyzers running software version 1.3.3 or 1.3.4, if this issue occurs, you will need to place a new order for the CC combination test, - Select Home - Sample List task indicator, then select Active in the upper left corner of the Sample List to display samples "In Progress". - Program the CC combination tests for these "In Progress" samples using a new rack and Sample ID, apply the new barcode, then load the rack onto the analyzer for processing. - Set aside the rack with "In Progress" samples and contact your Beckman Coulter representative to clear them. Firm recommends sharing the content of this letter with your laboratory and/or Medical Director. Firm has resolved this issue in Software version 1.3.5 and a firm service representative will contact you to schedule the software upgrade. Share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter. Complete and return the response form via email to: regaffairsfax@beckman.com If you have any questions regarding this product, please contact your local firm representative, or use the following link for a listing of local contact information. https://www.beckmancoulter.com/en/support/contact-us

Distribution Pattern Internationally distribution to the countries of Poland, India, Dominican Republic, Lebanon, Italy, Brazil, El Salvador, Saudi Arabia, Germany, France, Nepal, Canada, Chile, Spain, Malaysia, Switzerland, Morocco, Philippines, Curacao, Qatar, United Arab Emirates, Hungary, Czechia, Oman, New Zealand, Thailand, Indonesia, Algeria, United Kingdom of Great Britain and Northern Ireland, Namibia, Turkey, Republic of Korea

Code Information / Serial Numbers REF/UDI-DI: C63522/14987666545089, C63521/14987666545072. Software Versions: SW 1.3.3 and 1.3.4

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.