Medical Device Recall: Z-0028-2026

FDA Recalls dataset · Firm: NOXBOX LTD · Status: Open, Classified

Recall Action Class: N/A Date Initiated: 2025-09-09 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0028-2026.

Recall Res Number
Z-0028-2026
Recalling Firm / Manufacturer
FEI Number
3012796317
Recall Status
Open, Classified
Recall Classification
N/A
Date Initiated
2025-09-09
Date Posted
2025-10-06
Date Terminated
N/A
Product Code
MRN
Event Number
97517
Product Quantity
1667 units
Product Description NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I

Reason for Recall Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).

Action Details On September 9, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: - Use the instructions provided to troubleshoot the device. - Complete and return the attached response form to Linde Gas & Equipment as instructed on the form. Recalling firm actions: NOxBOX Ltd. is in the process of developing a new software version that will include a modification to the software to resolve this issue, which will be released pending required regulatory review and approval. Once the software version is released, NOxBOX Ltd. will coordinate with Linde Gas & Equipment in the U.S. to update your device. If you have any questions about this notice or need assistance, please reach out to your local Clinical Educator or the Customer Excellence Center at 1-833-NOX-VENT (669-8368) or at LG.US.NOXIVENT@Linde.com.

Distribution Pattern Domestic: TN

Code Information / Serial Numbers Model Numbers: (1) NOXBOXI (NBL (Manufacturer)), (2) REQNOXBOXI (Linde (U.S. Distributor)); UDI-DI: 05060541640009; All serial numbers.

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.