Medical Device Recall: Z-0027-2025

FDA Recalls dataset · Firm: Windstone Medical Packaging, Inc. · Status: Open, Classified

Recall Action Class: N/A Date Initiated: 2024-08-16 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0027-2025.

Recall Res Number
Z-0027-2025
Recalling Firm / Manufacturer
FEI Number
1000125955
Recall Status
Open, Classified
Recall Classification
N/A
Date Initiated
2024-08-16
Date Posted
2024-10-08
Date Terminated
N/A
Product Code
LRO
Event Number
95306
Product Quantity
642 kits
Product Description Aligned Medical Tonsil Tray REF AMS9300A that contains the Cardinal Health Salem Sump PVC tubes. Product packaged in a convenient manner for use in a general clinical procedure.

Reason for Recall Due to convenience kits containing a kit component that is under a recall action for improper use.

Action Details On 08/16/2024, the firm email an "URGENT: MEDICAL DEVICE PRODUCT CORRECTION" Letter to customers informing them that Cardinal Health is conducting a product correction for its Salem Sump PVC Tubes. These Sump PVC tubes were included in the Aligned Medical convenience kits. The correction for this Recall is an updated product label, an electronic instruction for use (eIFU). The updated labeling is to address improper use of the Sump PVC tubes which can lead to breakage of the Salem Sump Anti Reflux Valve (ARV) that increases risk to patients. Customers are instructed to: 1. Review your inventory for the affected kits, all lots are impacted 2. Communicate the change to the use instructions with all personnel that utilize Salem Sump" products contained within AMS kits. 3. If you have further distributed this product, please identify your distribution partners, and notify them at once of the product correction. Your notification to your customers should include a copy of this notification letter. 4. Please complete and return the AMS Urgent Medical Device Product Correction Response Required form within 5 business days and return via email to fieldcorrectiveaction@alignedmedical.com even if you have no affected product on hand. For questions or concerns, contact Aligned Medical Solutions at fieldcorrectiveaction@alignedmedicalsolutions.com

Distribution Pattern U.S. Nationwide distribution in the states of CA, MO, OH, VA, and WY.

Code Information / Serial Numbers Catalog Number: AMS9300A UDI-DI code: B098AMS9300A0 Lot Numbers: 168030 171362 172981 177913 179608 181537 185126 189873 189874 196460 196461 198850 200902 206846 208255 209391

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.