Medical Device Recall: Z-0025-2007

FDA Recalls dataset · Firm: Abbott Diagnostic International, Ltd. Carr 2 Km 58.0 Cruce Davila · Status: Terminated

Recall Action Class: N/A Date Initiated: 2006-09-07 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0025-2007.

Recall Res Number
Z-0025-2007
FEI Number
2623532
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2006-09-07
Date Posted
2006-10-12
Date Terminated
2011-09-09
Product Code
JIT
Event Number
36405
Product Quantity
624
Product Description Architect FSH Calibrators (LN 6C24-01), for in vitro diagnostic use.

Reason for Recall Following a customer complaint, Abbott found an atypical stability profile for the lot of the calibrators listed in this recall. The investigation to date has shown that both controls and patient results have shifted upwards over time together.

Action Details The Device Recall Letter dated September 07, 2006, and Customer Reply Form was sent via Federal Express Priority mail to all customers who received the affected lots.

Distribution Pattern Worldwide, including USA, Puerto Rico, Canada, Mexico, Guatemala, Costa Rica, Venezuela, Colombia, Ecuador, Chile, Uruguay, Argentina, Brazil, Dominican Republic, El Salvador, Germany, Japan, Singapore, Hong Kong, Taiwan, Thailand, South Korea, Australia, Honduras, Trinidad-Tobago, St. Vincent, Bahamas, Panama, and Cayman Islands.

Code Information / Serial Numbers Lot # 33363Q100 Expiration Date 11/04/2006

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.