Medical Device Recall: Z-0023-2010

FDA Recalls dataset · Firm: Hospira Inc · Status: Terminated

Recall Action Class: N/A Date Initiated: 2009-08-11 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0023-2010.

Recall Res Number
Z-0023-2010
Recalling Firm / Manufacturer
FEI Number
2921482
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2009-08-11
Date Posted
2009-10-06
Date Terminated
2010-11-30
Product Code
FRN
Event Number
52866
Product Quantity
204,999 total
Product Description Plum A+3, list #s 12348-04-03, 12348-04-05, 12348-04-07, 12348-04-51; AC cord manufactured by Electri-Cord, Westfield PA, Infusion pumps manufactured by Hospira, Morgan Hill, CA.

Reason for Recall Fire/Shock hazard-- A component used in the device may cause sparks/flashes and poses a fire and shock hazard.

Action Details Firm initiated its recall on 08/14/2009. A nationwide press release was issued to AP, and an Urgent: Device Field Correction letter to consignees was sent via UPS. The letter recommended that customers check their infusion pumps and spare parts in their inventory to determine if the specified AC Power Cords are in use at their facility; and, to inspect the affected cords for bent or cracked prongs and bridge connection, burnt plastic, or excessive wear and tear. Also, it was recommended that customers immediately discontinue use of the affected cords exhibiting the described characteristics and contact their local Hospira representative for assistance. The letter stated that customers may continue to use and monitor cords that are not exhibiting the described characteristics until replacement cords are sent to their facility. Hospira will contact customers to schedule replacement. Customers are to complete and return the attached Reply Form. Question should be directed to Hospira Customer Service at 1-800-241-4002.

Distribution Pattern Worldwide Distribution

Code Information / Serial Numbers AC power cords Part Numbers: 826-85559-001,  826-85559-003,  826-85559-008,  826-85559-010, and  826-96057-003.

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.