Medical Device Recall: Z-0020-2016

FDA Recalls dataset · Firm: Siemens Medical Solutions USA, Inc · Status: Terminated

Recall Action Class: N/A Date Initiated: 2015-08-14 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0020-2016.

Recall Res Number
Z-0020-2016
Recalling Firm / Manufacturer
FEI Number
2240869
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2015-08-14
Date Posted
2015-10-05
Date Terminated
2016-12-05
Product Code
JAK
Event Number
72091
Product Quantity
113 total
Product Description SOMATOM Definition Flash; intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles.

Reason for Recall software bug issues for SW-Version VA48A_SP0. The following safety issues were resolved: 1) Correction to improve visual warning and error indication son the gantry display. 2) Correction to improve acquisition data in order to optimize image quality. 3) Correction to improve robustness and general system behavior in some exception handling procedures. 4) Correction to improve auto post processin

Action Details A customer advisory notice, dated 8/14/15, was sent to direct accounts advising them of the issue to observe the safety notice and comply with the corresponding measure until the update has been fully completed.

Distribution Pattern Nationwide Distribution.

Code Information / Serial Numbers Model# 10430603 with Serial numbers 73476 73586 73544 73648 73636 73621 73533 73423 73424 73520 73493 73459 74191 73487 73492 73494 73497 73013 73657 73535 73667 73490 73513 74192 73697 73635 73471 73405 73600 73558 73591 73546 73478 73504 73413 73674 73445 73547 73420 73430

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.