Medical Device Recall: Z-0016-2017

FDA Recalls dataset · Firm: Panoramic Rental Corp. · Status: Terminated

Recall Action Class: N/A Date Initiated: 2016-08-19 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0016-2017.

Recall Res Number
Z-0016-2017
Recalling Firm / Manufacturer
FEI Number
1832462
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2016-08-19
Date Posted
N/A
Date Terminated
2018-01-11
Product Code
EHD
Event Number
75017
Product Quantity
13,340
Product Description Panoramic X-ray Model PC-1000, Laser-1000 packed in a crate Product Usage: The PC-1000 will enable the user to take panoramic x-ray images. The PC-1000/Laser1000 will enable the user to take panoramic x-ray images as well as cephalometric x-ray images.

Reason for Recall Panoramic Corporation is performing a voluntary Medical Device Correction on certain PC-1000 X-ray systems due to the potential for a safety nut to be ineffective, which could result in blunt force trauma that may impact the head, neck or shoulders.

Action Details Panoramic Corporation sent an Urgent Medical Device Correction letter dated August 19, 2016 to customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the attached form and return to Panoramic. For questions contact Panoramic at correction@pancorp.com or 888-381-2577 Monday through Thursday 8:00 AM to 6:00 PM and Friday 8:00 AM to 5:00 PM EST in order to verify the effectiveness of the safety component.

Distribution Pattern US Nationwide Distribution including Puerto Rico

Code Information / Serial Numbers Model Number(s): 800724-1, 800724-3, Serial numbers 6001 to 19340

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.