Medical Device Recall: Z-0015-2012

FDA Recalls dataset · Firm: Deroyal Industries, Inc. Lafollette · Status: Terminated

Recall Action Class: N/A Date Initiated: 2011-08-23 Recall record compiled from public FDA source records.

Recall Parameters & Description

Normalized public record details for FDA recall code Z-0015-2012.

Recall Res Number
Z-0015-2012
Recalling Firm / Manufacturer
FEI Number
3005011024
Recall Status
Terminated
Recall Classification
N/A
Date Initiated
2011-08-23
Date Posted
2011-10-06
Date Terminated
2013-01-30
Product Code
OGD
Event Number
59854
Product Quantity
940 units
Product Description Custom anesthesia kits labeled as follows: (1) DeRoyal (R) Adult Exp Anesthesia Circuit, REF 86-001737, Rx Only, NON-STERILE, Manufactured by: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA; Lot Number 25660996 (2) DeRoyal (R) Adult Anesthesia Circuit, REF 86-001781, Rx Only, NON-STERILE, Manufactured by: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA; Lot Numbers: 25574115, 25624514, 25661260, 25702578, 25766272, 25835901, 25835910, 25873406, 25906061, 25960681 (3) DeRoyal (R) Latex Safe Anesthesia Kit, REF 86-001641, Rx Only, NON-STERILE, Manufacturer: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA: Lot Numbers: 25573171, 25765667, 25834853, 25834861 (4) DeRoyal (R) Custom Anesthesia Kit, REF 86-001619, Rx Only, NON-STERILE, Manufacturer: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA: Lot Number 25573155 (5) DeRoyal (R) Anesthesia Kit, REF 86-001669, Rx Only, NON-STERILE, Manufacturer: DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849, USA: Lot Numbers: 25313484, 25702324, 25977977 Intended Use: Varies by kit; an anesthesia breathing circuit is a device that is intended to administer medical gases to a patient during anesthesia. It provides both an inhalation and exhalation route and may include a connector, adaptor, and Y-piece.

Reason for Recall The firm manufactured kits containing a HCH device, manufactured by Vital Signs, that was subsequently recalled because the product may have an occlusion that could prevent proper flow of medical gases or oxygen to the patient.

Action Details The firm, DeRoyal, sent a "URGENT: MEDICAL DEVICE RECALL" letter dated September 13, 2011 to its customers. The letter describes product, problem, and actions to be taken. Customers are instructed to identify inventory being recalled and place in quarantine to prevent further use. Constomers were asked to return affected product they have in inventory. DeRoyal will replace this product free of charge. Customers were asked to complete the attached Recall Response Spreadsheet and return as per the instructions. Constomers were asked to return affected product they have in inventory. DeRoyal will replace this product free of charge. Distributors were instructed to perform a sub-recall to their customers. Customers with questions should call 865-362-1037 between 9am and 5pm EST or email recalls@deroyal.com.

Distribution Pattern Nationawide Distribution USA - including the states of: CA, FL, GA, NJ, and WA.

Code Information / Serial Numbers REF 86-001737, Lot Number 25660996 REF 86-001781, Lot Numbers: 25574115, 25624514, 25661260, 25702578, 25766272, 25835901, 25835910, 25873406, 25906061, 25960681 REF 86-001641, Lot Numbers: 25573171, 25765667, 25834853, 25834861 REF 86-001619, Lot Number 25573155 REF 86-001669, Lot Numbers: 25313484, 25702324, 25977977

Related Establishments / Registrations

Public Record Disclaimer

This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.