Product Description
Ammonia Liquid Stable Reagent, Enzymatic Method. Catalog number TR60101. in vitro diagnostic.
Reason for Recall
Method is susceptible to positive interference from endogenous ALT in patient samples.
Action Details
The firm has updated the Limitations section of the Ammonia Reagent Package Insert and circulated a Technical Bulletin dated July 24, 2007 to all customers that have purchased Ammonia Reagent that is still within expiration date. A Notification of Change has been introduced and shall be placed within subsequent Ammonia Reagent Kits with a copy of the updated package insert; the notification of change informs customers that a change has been made to the package insert. Notifications of Change are placed within kits for 6 months from the date of issue.
Distribution Pattern
Worldwide, including USA, Ecuador, and Canada.
Code Information / Serial Numbers
Lot numbers FL029 exp 8/30/07, GB059 exp 11/30/07, GD005 exp 12/30/07, GE093 exp 1/30/08, GH034 exp 4/30/08, GJ039 exp 7/30/08, HB064 exp 11/30/08, and HD035 exp 1/30/09.
Public Record Disclaimer
This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.