Product Description
Mouse anti-TTF-1 concentrate used in a panel designed to differentiate between different types of cancer when used to test a patient.
Reason for Recall
Mouse anti-TTF-1 had become contaminated by immunoglobulins from mouse serum that was co-eluted during purification leading to false readings.
Action Details
The firm has conducted telephone calls and visits to its consignees, with confirmation done by phone and fax. The firm has recovered or accounted for 100% of the products, either returned or destroyed.
Distribution Pattern
The Firm distributed to a total of 8 domestic consignees and an unspecified number of foreign consignees. The products were distributed to MD, GA, TX, OH, IL, DC, NY and CA. Foreign countries distributed to include France, Portugal, China, Italy, Israel, Philippines, and Germany.
Code Information / Serial Numbers
Product Number 18-0221, no 510(k) number, Product Code MXZ, lot code 40280262.
Public Record Disclaimer
This page represents an index of public-record FDA device recalls. It does not imply current safety or compliance status of the product. AvaSpecs is a private aggregator and is not affiliated with the FDA or any government agency. This page does not constitute medical, safety, legal, regulatory, or compliance advice. Refer to official government sources for primary safety alerts.